S012: Test Step 1 - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 5
Selecting test step 1
Checking the module
59 38 399 D3352
D3352.076.01.26.02 09.2016
S012.2: Inquiry of the CAN status register of the modules
5.10.1

S012: Test step 1

Presence display of modules
Selection field Parameter/Display
1
Subunit
Counter value
2
of CAN bus events
Presence code behind
the counter value:
P = Module is "present"
L = Module is "lost"
1.
Select test step 1 in selection field 2 with the arrow keys and confirm your
selection by pressing the service key
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
-
+
S012
+
-
1
2
3
2.
Select the desired module using the arrow keys of selection field 1.
The counter value of the CAN bus events processed so far (since the last
switch-on of the unit) of the selected module is displayed in selection field
2 with the presence code of the module ("L" or "P") (see table).
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
-
+
DX61
1234P
2
3
T
After the required module has been selected the test key
Plus
ready
(XG
/ 3D
) or the LED above the test key
5.10 Service routine S012
Range of values
DX1 - DX88
.
XG 5 / 3 / 3 PPE: Multipad
S12
S012_______1____
1
XG 5 / 3 / 3 PPE: Multipad
DX61___OP________
1
(XG 5 / 3 / 3 PPE).
T
2
3
2
3
T
5 – 65

Hide quick links:

Advertisement

Table of Contents
loading