4.3 Adjusting the panoramic X-ray unit
1.
4 – 10
Dismantling the fixed diaphragm
4.3.2
(XG 3 / 3 PPE only)
The fixed diaphragm must be dismantled in order to perform the sensor
adjustment of the ORTHOPHOS XG 3 / 3 PPE X-ray unit. To do this, proceed
as follows:
1.
Remove the covers:
– Tube assembly, front
– Tube assembly, rear
Tab 4
59 38 399 D3352