Tab 3
Check operating status of motors.
Visual inspection:
• Is the motor intact?
• Are the plug connections OK?
yes
Temporarily install replacement motor
and check unit function.
• Is the unit functioning properly?
yes
Replace motor, see section 6.3 or
section 6.4
59 38 399 D3352
D3352.076.01.26.02 09.2016
3.4
Checking the motors
Replace the motor and/or fix the plug
No
connections.
Motor is OK!
No
Continue troubleshooting according to
error list (see section 2.5).
3.4 Checking the motors