Reading Out Sorting Judgment Result - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Sorting Based on Measured Results (Comparator Function)

Reading out Sorting Judgment Result

Reading out Sorting Judgment Result
The sorting judgment is obtained through the comparator function according
to the flow shown in
Figure 7-4
result.
Figure 7-4
Relationship between display output and comparator sorting result
If an overload is detected, the sorting judgment cannot be performed and
therefore ---- is displayed in the upper row. If Low C reject is detected,
normal sorting judgment is performed but LOWC is displayed in the upper
row instead of the sorting judgment result (OUT, AUX, or BIN1 to BIN9).
The sorting judgment result can be read out through the SCPI command.
132
Figure
7-5.
shows the relationship between the screen display and the sorting
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter