Sorting Based on Measured Results (Comparator Function)

Reading out Sort Count of Each BIN (BIN count function)

Reading out Sort Count of Each BIN (BIN count function)
You can count the number of DUTs sorted into each BIN by turning ON the BIN
count function. The maximum value of the count is 999999. If this value is
exceeded, the count will not increase but remain at 999999 (it does not return
to 0).
When the MULTI correction function is ON, channel-by-channel count is
performed separately from the usual count (total of all channels). You can read
out the result by using the SCPI command.
You can read all the BIN count values with the
:CALCulate1:COMParator:COUNt:DATA? command. For more information,
refer to the Programming Manual.
The overload count value can be read only through the SCPI command and
cannot be displayed on the screen.

Setup procedure

The BIN count value is displayed in the BIN COUNT DISPLAY page.
The setup procedure is given below.
Step 1.
Press [Display Format].
Step 2.
Press the BIN COUNT softkey.
Step 3.
Use the following softkeys:
Softkey
COUNT ON
COUNT OFF
RESET COUNT
Clearing the BIN count values initializes all of the count values to 0.
The BIN count value for each channel when the MULTI correction function
is ON can be checked only through SCPI command and cannot be
displayed on the screen.
The overload count value of each channel when the MULTI correction
function is ON, can be read with the
:CALCulate1:COMParator:COUNT:MULTi:OVLD? command. For more
information, refer to the Programming Manual.
134
Function
Turns ON the count function.
Turns OFF the count function.
Resets the count function.