Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions
Turning on OFFSET correction
Use the following procedure to set the OFFSET correction to ON/OFF.
Step 1.
Press [Meas Setup].
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the OFFSET field.
Step 4.
Press the ON softkey to turn ON the OFFSET correction function.
If you change the setup of the measurement parameter with the OFFSET
correction ON, the OFFSET correction is automatically set to OFF.
You cannot separately set the ON and OFF states as the primary parameter
and secondary parameter. However, if you set the correction value to 0, the
state is effectively the same as OFF even if the OFFSET correction is ON.
Therefore, you can set up separate ON/OFF states by setting the correction
value of either parameter to 0.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter