Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
USB memory notes
Attention should be given to the following points when using a USB memory
device with the E4981A
•
•
•
•
•
Saving/Recalling instrument configuration states
Overview of instrument configurations
Pressing the [Save/Recall] opens the CATALOG page. You can save/recall
instrument configuration states in either of two ways:
•
•
You can save up to 10 registers into the internal memory and into a single USB
memory device. Also, register numbers have extensions as shown in
Table 4-9
Extensions for register numbers
Internal memory
In this page, you can configure each of the following controls with the cursor
placed in the corresponding field (denoted in parentheses).
•
•
78
Use a USB memory whose interface is USB 1.1.
Use a USB memory that is USB mass storage class compliant and
formatted with FAT16 or FAT32.
Use the USB memory solely for the E4981A. Otherwise, other data
previously saved in the USB memory could be erased.
If you cannot save into or recall from the USB memory, use another USB
memory device.
Keysight Technologies shall not be responsible for nor assume any liability
for data loss in your USB memory device after using it with the E4981A.
Save into the internal memory
Save into USB memory
Register number
Classification
(No. field)
2 through 8
USB memory
10 through 19
Medium mode (MEDIA field)
Register number (No. field)
Extension
0
Recalled when the [Recall A] key is pressed.
1
Recalled when the [Recall B] key is pressed.
-
9
Auto recall
-
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Table
4-9: