Using Handler Interface

Timing Chart

Timing Chart
Figure 8-3
the figure indicates that the E4981A is processing data after analog
measurement and the output signals are invalid. For details on measurement
time, refer to
Figure 8-3
Timing chart for handler interface
e4981aue0039
Table 8-3
Values of T1–T5
T1
Trigger pulse width
Trigger response time of /READY_FOR_TRIG, /INDEX
T2
and /EOM
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
shows the timing chart. The section where /Data is unpredictable in
Table 8-3 on page
T1
E X T _T R IG
/IN D E X
/EOM
/READY FOR
_TRIG
/Data
Previous sorting result
145.
T3
T2
/Data :/BIN1
/BIN9,/OUT_OF_BINS,
/OVLD /LOWC_OR_NC
MEAS TIME
NA
NA
T4
T5
Unpredictable
Sorting result
Minimum
Typical
value
value
1 µs
40 µs
145