Information for Replacing 4268A, 4288A with E4981A

Functional Comparison between 4268A, 4288A, and E4981A

Functional Comparison between 4268A, 4288A, and E4981A
Table B-3
Table B-3
Functional comparison between 4268A, 4288A, and E4981A
Function
Measurement
Reset
condition
Measurement parameter setup
Measureme
nt signal
setup
Measurement range mode
Measureme
nt time
(integration
time)
Averaging count
Cable length setup
Source delay setup
218
compares the functions of the 4268A, 4288A, and E4981A.
E4981A
Can be executed
using the front panel
or SCPI command.
Selectable from
Cp-D, Cp-Q,
Cp-G,Cp-Rp, Cs-D,
Cs-Q, or Cs-Rs.
Frequency
Selectable from 120
Hz, 1 kHz or 1 MHz.
1 MHz frequency shift
Can be changed to
+2%, –2%, +1%,
–1%, or 0% .
OSC Level
Can be set up within
the range of 0.1 to
1 V in steps of
0.01 V.
Level monitor function
AC Level monitor is
always ON.
Signal level compensation
Available at 120 Hz
(SLC)
or 1 kHz.
Output mode
Continuous,
Synchronous.
Selectable from the
auto range or hold
range.
Mode
Selectable from
Short, Medium, or
Long (Using SCPI
command only).
Time
Selectable from 1, 2,
4, 6, or 8.
Can be set up freely
within the range of 1
to 256.
Selectable from 0 m,
1 m, or 2 m.
Can be set within
0 - 1 sec in steps of
100 µs.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
4268A
4288A
Can be executed
Can be executed
using the front panel
using the front panel
or GPIB command.
or GPIB command.
Selectable from
Selectable from
Cp-D, Cp-Q,
Cp-D, Cp-Q, Cp-G,
Cp-G,Cp-Rp, Cs-D,
Cp-Rp, Cs-D, Cs-Q,
Cs-Q, or Cs-Rs.
or Cs-Rs.
Selectable from 120
Selectable from
Hz or 1 kHz.
1 kHz or 1 MHz.
N/A
Can be changed to
+2%, +1%, or –1% .
Can be set up within
Can be set up within
the range of 0.1 to
the range of 0.1 to
1 V in steps of 0.01
1 V in steps of 0.1 V.
V.
Voltage value and
Voltage value and
current value can be
current value can be
monitored.
monitored.
ALC is available
N/A
Continuous,
N/A
Synchronous.
Selectable from the
Selectable from the
auto range or hold
auto range or hold
range.
range.
Selectable from
Selectable from
Short, Medium, or
Short or Long.
Long.
N/A
N/A
Can be set up freely
Can be set up freely
within the range of 1
within the range of 1
to 256.
to 256.
Selectable from 0 m,
Selectable from 0 m,
1 m, or 2 m.
1 m, or 2 m.
Can be set within
N/A
0 - 1 sec in steps of
1 µs.