Turning On Short Correction - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions

Turning on SHORT correction

Use one of the following two ways to set the SHORT correction to ON.
Measuring data for SHORT correction
Execute measurement of the data for the SHORT correction. When the
measurement finishes successfully, the SHORT correction is automatically set
to ON. For information about the measurement procedure for the data used in
the SHORT correction, refer to
page
Setting up ON/OFF directly
Use the following procedure to set the SHORT correction to ON/OFF.
Step 1.
Press [Meas Setup].
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the SHORT field.
Step 4.
Press the ON softkey to turn ON the SHORT correction function.
If you change the setup of the cable length or frequency shift (1 MHz) with
the SHORT correction ON, the warning message "Need corr meas" appears
and the SHORT correction is automatically set to OFF. Along with SHORT
correction, OPEN and LOAD correction is set to OFF as well.
If you set the SHORT correction to ON using the above procedure when the
setup of the cable length or frequency shift differs from that during
measurement of the data for SHORT correction, the warning message
"Need short meas" appears. Even if this warning message appears, the
SHORT correction is set to ON. However, you must again measure the data
for SHORT correction to ensure accurate measurement.
When the measurement result becomes overload, "Measurement failed"
error is displayed and SHORT correction data is not updated.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
47.
"Measuring data for SHORT correction" on