Setting Up Limit Ranges - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
Figure 7-2
Example of limit range settings (case 2)
Table 7-2
Lower and upper limit values of limit ranges for
modes)

Setting up limit ranges

Notes on setup
126
Absolute mode
Lower
limit
value
BIN1
0.99 nF
1.01 nF
BIN2
0.98 nF
1.02 nF
BIN3
0.97 nF
1.03 nF
If you set up the upper limit value to a value equal to or less than the lower
limit value, the limit range is not used. This operation is equivalent to
setting the limit range to OFF.
If BINs overlap, the resulting measurement is sorted into the BIN of the
smallest number (refer to
page
133). Therefore, you need to set up the limit ranges starting from the
narrowest to the widest as shown in
For the tolerance mode, the reference value is not required to be within the
limit range (between the lower limit value and upper limit value).
Gaps between the limit ranges are allowed. Therefore, you can set up the
limit ranges as shown in
Tolerance mode (reference value: 1 nF)
Absolute
Upper
Lower
limit
limit
value
value
–0.01 nF
0.01 nF
–0.02 nF
0.02 nF
–0.03 nF
0.03 nF
Figure 7-5, "Sorting judgment flow," on
Figure
7-2.
Figure
7-3.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Figure 7-2
(comparison of
Percent
Upper
Lower
limit
limit
value
value
–1%
–2%
–3%
Upper
limit
value
1%
2%
3%

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a-001E4981a-002

Table of Contents