Selecting Single/Multiple Correction Mode - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
Description of softkeys
To enable/disable or otherwise control the behavior of load correction, use the
following softkeys:
Softkey
ON
OFF
MEAS LOAD

Selecting single/multiple correction mode

Functional description
E4981A can store up to 256 sets of OPEN/SHORT/LOAD correction data. In
addition, it can store one set of the standard's reference value data at a
specified frequency point. In multiple correction mode, you can switch among
up to 256 data sets to carry out correction.
The correction features default to single correction mode.
In multiple correction mode, the channel number for which correction data is
selected.
To set up single/multiple correction mode:
Step 1.
Press [Meas Setup].
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the MULTI field.
Step 4.
Use the following softkeys:
Softkey
ON
OFF
For more information on measurement parameters, see
measurement parameters" on page
Step 5.
Use the cursor keys to select the CH field.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Description
Enables load correction.
Disables load correction.
Starts load correction.
Description
Turns ON the multiple correction mode.
Turns OFF the multiple correction mode.
"Setting up
40.
107