Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)

Using the auto recall feature

Functional description
You can have the E4981A, at start-up, automatically recall the configuration
states previously saved in register number 9 on Internal memory.
To use the auto recall feature:
Step 1.
Press [Save/Recall].
Step 2.
Use the cursor keys to select the No. 9 [A] field.
Step 3.
Press the RECALL softkey, to configure the instrument using the instrument
configuration information previously saved in register 9 on the Internal
memory.
Transferring auto recall setting to another E4981A
Using external USB memory
Step 1.
Perform auto recall.
Step 2.
Connect external USB memory.
Step 3.
Press [Save/Recall].
Step 4.
Use the cursor keys to select MEDIA to EXT.
Step 5.
Use the cursor keys to select memory location from 10 to 19.
Step 6.
Press the SAVE softkey to save the data in the USB memory.
Step 7.
Connect the USB memory to another E4981A, recall the saved state.
Step 8.
After recalling the state, save it to internal memory No. 9[A].
Using *LRN? command
Step 1.
Perform auto recall.
Step 2.
Retrieve the instrument configuration using *LRN? command via
GPIB/LAN/USB.
Step 3.
Send the instrument configurations (retrieved in Step 2) to another E4981A.
Step 4.
Save the data to internal memory No. 9[A].
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
85