Getting Started
Learning Basic Measurement Procedure

Setting up basic measurement conditions

This section describes how to set up basic measurement conditions for
capacitor measurement. For an actual measurement (for example,
measurement that provides better accuracy or measurement based on
measurement time), you need to set up measurement conditions that are
appropriate to the task, not necessarily those described here.
Initializing the Instrument
This section describes how to return the settings of the E4981A to their preset
values.
Step 1.
Press the [Preset] key.
Step 2.
Use the softkeys to select one of the following four default states:
Table 3-1
Four default states of E4981A and how it is initialized
Default state
CLEAR SETTING
CLEAR SET&CORR When you initialize the instrument into this state,
FACTORY DEFAULT When you initialize the instrument into this state, it
LAN RESET
For more information on each default state and affected settings, see
Appendix C , "Initial Settings," on page
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
How the instrument is initialized
When you initialize the instrument into this state, all
basic parameters configurable through the front panel
and SCPI commands are cleared. (You get the same
result by issuing the :SYST:PRES command).
calibration data listed in the initial setting list are all
cleared.
*RST command).
reverts to factory default settings with all
user-configurable data cleared.
When you initialize the instrument into this state, the
LAN setting is returned to the factory default state.
1. It takes a few seconds for the initialization to complete.
1
(You can get the same result by issuing the
227.
39