Using Scanner Interface
Using Multi-correction Function
The procedure to turn ON/OFF the multi-correction function is given below.
Step 1.
Press [Meas Setup].
Step 2.
Press CORRECTION softkey.
Step 3.
Use the cursor keys to select the MULTI field.
Step 4.
Use the following softkeys:
Softkey
ON
OFF

Selecting a channel

Making a selection using the front panel
The procedure to select a channel using the front panel keys is given below.
Step 1.
Press [Meas Setup].
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the CH field.
Step 4.
Use the softkeys or numeric entry keys to enter the channel number. When
data is entered using the numeric keys, the softkeys change to units labels (x1).
Softkey
INCR++
INCR+
DECR-
DECR--
154
Function
Turns ON the multi-correction function.
Turns OFF the multi-correction function.
Function
Increments the channel number in steps of 10.
Increments the channel number in steps of 1.
Decrements the channel number in steps of 1.
Decrements the channel number in steps of 10.