Checking Displaying/Setting Up Data For Load Correction - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Preparation for Accurate Measurement (Executing Correction)
Checking Displaying/Setting Up Correction Data

Checking displaying/setting up data for LOAD correction

Displaying data/Selecting parameter format for LOAD correction
The parameter format of LOAD correction data depends on the parameter
format of the standard value for the LOAD correction and cannot be selected
independently.
Refer to
value)" on page
Setting up data for LOAD correction
You can setup the data for LOAD correction by the following procedure.
Step 1.
Press the [Meas Setup] key.
Step 2.
Press the CORRECTION softkey.
Step 3.
Use the cursor keys to select the LOAD Cp-D field.
Step 4.
Select Cp- ... or Cs- ... which you required.
Step 5.
Use the right cursor keys to select the A field.
Step 6.
Input your desired value for the primary parameter for LOAD.
Step 7.
Use the right cursor keys to select the B field.
Step 8.
Input your desired value for the secondary parameter for LOAD.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
"Defining standard for LOAD correction (setting up LOAD reference
103.
111