Setting The Trigger Mode; Perform Successive Measurements Automatically; Specifying Measurement Timing - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Executing Measurement
Starting (Triggering) Measurement

Setting the trigger mode

Step 1.
Press [Meas Setup].
Step 2.
Use the cursor keys to select the TRIG field.
Step 3.
Select the desired trigger mode by pressing the appropriate softkey
Softkey
INT
MAN
EXT
BUS

Perform successive measurements automatically

Select the internal trigger mode according to the procedure in
trigger mode" on page
repeated, triggered by the internal trigger signal.

Specifying measurement timing

Generating a trigger manually
Step 1.
Select the manual trigger mode according to the procedure in
trigger mode" on page
Step 2.
Press the [Display Format] key.
Step 3.
Press the [Trigger] key to make a single measurement.
116
Function
Puts the instrument into internal trigger (INT) mode.
Puts the instrument into manual trigger (MAN) mode.
Puts the instrument into external trigger (EXT) mode.
Puts the instrument into bus trigger (GPIB/USB/LAN) mode.
116. In this mode, measurements are automatically
116.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter