Selecting Single/Multiple Correction Mode - Keysight E4980AL User Manual

Precision lcr meter
Hide thumbs Also See for E4980AL:
Table of Contents

Advertisement

Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page

Selecting Single/Multiple Correction Mode

Functional Description
When equipped with Option 301 (Scanner Interface), E4980A/AL can store up
to 128 sets of open/short/load correction data. In addition, it can store one set
of the standard's reference value data at a specified frequency point. In
multiple correction mode, you can switch among up to 128 data sets to carry
out correction. Note that the correction features default to single correction
mode.
When you use multiple correction mode, you cannot perform open/short
correction by the interpolation method (open/short correction at particular
frequency points) but must perform correction at specified frequency
points.
User-specified frequency points can be configured separately between
single and multiple correction modes.
In multiple correction mode, the channel number for which correction data are
selected is displayed on the right-hand side of the CH indicator.
To set up single/multiple correction mode:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the MODE field.
Step 4. Use the following softkeys:
Softkey
SINGLE
MULTI
For more information on measurement parameters, see
Function" on page
When Option 301 (Scanner Interface) is not installed and you attempt to
switch to multiple correction mode, a Scanner I/F not installed error
message appears in the system message area.
132
Description
Turns ON single correction mode.
Turns ON multiple correction mode.
47.