Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company’s sole decision. Information in this publication replaces all previously released materials.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
Page 5
Do Not Operate With Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
RIGOL Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data.
System denotes pressing Utility on the front panel and then pressing System. User Manuals of this Product: The user manuals of this product include the Quick Guide, User Guide, Programming Guide, Data sheet and etc. For the latest version of the desired manual, download it from www.rigol.com. DS1000Z-E Service Guide...
RIGOL Contents Contents Guaranty and Declaration ................. I Safety Requirement ................. II General Safety Summary................II Safety Notices and Symbols ..............IV Document Overview .................V Chapter 1 Disassembly and Assembly ..........1-1 Notices to Disassembly and Assembly ............1-1 Exploded View Drawing ................1-2 Disassemble the Rear Panel ..............
RIGOL Chapter 1 Disassembly and Assembly Chapter 1 Disassembly and Assembly Notices to Disassembly and Assembly Notices: Do not disassemble the instrument unless for working requirement. Only authorized personnel are allowed to disassemble the instrument. Cut off the power supply before disassembling the instrument.
RIGOL Chapter 1 Disassembly and Assembly Exploded View Drawing You need to get a basic understanding of the main parts of the instrument before disassembling and assembling the instrument. When disassembling or assembling the instrument, please follow the procedures and avoid scratching the part surface.
Page 11
RIGOL Chapter 1 Disassembly and Assembly Table 1-1 Replacement Parts List Part No. Qty. Name/Description 2010001921 Front Panel 2010000298 Large Knob 2010000315 Small Knob 2010003393 1020002199 Rubber Keypad 1020000579 M3x6 Torx Flat Head Countersunk Screw 1020001117 LCD Shield 2010002201 PCB Board of Rubber Keypad...
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Rear Panel Handle Rear Panel ② ② ① Figure 1-2 Disassemble the Rear Panel Part Description: ① 2 screws (M3x10 torx pan head screw, used for fastening the handle) ② 2 screws (M3x8 torx pan head self-tapping screw, used for fastening the...
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Metal Rear Panel ① ① ① ① ① ① ① Metal Rear Panel BNC Nut Figure 1-3 Disassemble the Metal Rear Panel Part Description: ① 11 screws (M3x6 torx flat head countersunk screw, used for fastening the metal front panel and the metal rear panel);...
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Knob ① ① ① ① ① ① Figure 1-4 Disassemble the Knob Part Description: ① 6 knobs Disassembly Step: Remove the 6 knobs (①) from the front panel. Tip: It is recommended that you use a lever-like tool to pry out the knob when disassembling the knob and place a soft pad at the force bearing point to avoid damaging the key film and knob.
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Front Panel ① ① ① ① ① ① ① Figure 1-5 Disassemble the Front Panel Part Description: ① 7 screws (M3x8 torx pan head self-tapping screw, used for fastening the front panel) Disassembly Steps: Use the T10 screwdriver to remove 7 screws (①) that secure the front panel.
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Keypad Board and LCD ① ③ Left Keyboard ③ ② ② ① Main Keyboard ③ ② ③ ② ② Figure 1-6 Disassemble the Keypad Board and LCD Part Description: ① 2 screws (M3x6 torx flat head countersunk screw, used for fastening the left keyboard) ②...
Page 17
RIGOL Chapter 1 Disassembly and Assembly Tip: Before disconnecting the keyboard cable and screen cable from the main board, pay attention to the connecting positions of the cables to avoid incorrect connection or incomplete connection when assembling the cables.
RIGOL Chapter 1 Disassembly and Assembly Disassemble the Main Board ① Figure 1-7 Disassemble the Main Board Part Description: ① 10 screws (M3x6 torx pan head combination screw, used for fastening the main board to the metal front panel); please remove them according to their actual positions).
RIGOL Chapter 1 Disassembly and Assembly Assembly Procedures The assembly procedures are simply the reversal of the disassembly procedures. Check whether the cables are correctly connected and whether all the screws are installed after completing each assembly procedure. DS1000Z-E Service Guide...
The commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem persists, please contact RIGOL and provide your device information (acquisition method: Utility System System Info).
Page 22
RIGOL Chapter 2 Troubleshooting&Maintenance Note: Using AUTO could automatically finish the above setting. 6. The display of waveform is ladder-like. (1) The horizontal time base might be too low. Increase the horizontal time base to increase the horizontal resolution and improve the display.
If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product. To get repair service, please contact with your nearest RIGOL sales or service office. There is no other warranty, expressed or implied, except such as is expressly set forth herein or other applicable warranty card.
RIGOL Chapter 2 Troubleshooting&Maintenance Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. Disconnect the instrument from all power sources.