Calibration User Configuration - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

9)
Repeat steps g) through h), to complete all the calibration standard Measures. Click [OK] to close the Measure
Mechanical Standard dialog box, to complete calibration.
By front panel keys
1)
Press 【Cal】in RESPONSE area, press the soft-key corresponding to [Cal Kit] to display the sub soft-key menu, and
press the soft-key corresponding to the required calibration kit model to select calibration kit.
2)
Press the soft-key corresponding to [Mechanical Calibration], and select calibration type in the displayed sub soft-key
menu.
3)
Conduct a calibration Measure by taking Thru Response calibration Measure for example. Press the soft-key
corresponding to [Response], and press the soft-key corresponding to [Thru Response] in the sub soft-key menu. Press
the soft-key corresponding to [FWD THRU]. After this menu item turns into [FWD THRU], a calibration Measure is
completed by pressing [DONE Response].
4)
Repeat steps b) through c), to complete the required calibration standard Measures.

5.2.2. Calibration User Configuration

Calibration User Configuration includes two settings, Save Cal File named as date and Do not save files when calibration finished.
Mouse or touch screen procedure
5.2.2.1.
1)
Single click [UTILITY], click [Configure] in the [System] menu, and click [User Config] in the submenu to display the User
Config dialog box.
2)
Tick Save Cal File named as date in the dialog box.
3)
Tick Do not save files when calibration finished in the dialog box.
Figure 5-8 Connect Calibration Standard Dialog Box
Figure 5-9 User Config dialog box
80