Calibration Setup; Starting Calibration - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

Measure requiring 12 errors correction to improve the Measure precision.
Measure requiring the maximum calibration precision.

5.2.Calibration Setup

Calibration can be set up and conducted by using the Calibration menu of th Analyzer.

5.2.1. Starting Calibration

By mouse or touch screen
1)
Click [Response], and click [Calibration ...] on the [Cal] submenu, to display the Calibration dialog box.
2)
If it is required to set the frequency range of calibration, click [Calibration Frequency] button in Calibration dialog box, to
display the Frequency Start/Stop dialog box. Click [OK] to close Frequency Start/Stop dialog box after completing
frequency setting.
3)
Click [Calibration Type] in Calibration dialog box, to display the Calibration Type dialog box, and Click [OK] to close
Calibration Type dialog box after completing calibration type selection.
 The selectable calibration type is relevant to the Measure parameters of the current Stimulus channel. For instance,
when Measure parameter is S11, Through Response as well as Through Response and Isolation Calibration can not be
selected.
 When conducting full dual-port calibration, if isolation calibration is not required, tick the [Omit Isolation (2-Port
Types)] check box.
Figure 5-1 Starting Calibration Wizard
Figure 5-2 Setting Calibration Frequency