By Front Panel Keys-Time Domain Transform Dialog Box - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

6)
Set filtering function of time domain gate: click [Gate Setting] option button in Time Domain Transform dialog box and click
to select [Gating] check box to enable filtering function of time domain gate.
 Click [Start], [Stop] or [Center], [Span] box to set gate border.
 Select gate type in [Gate Type] box: band pass, band stop.
 Select gate form in [Gate Shape]: minimum, normal, wide, maximum.
7)
If it is necessary to observe frequency domain response of the measured piece when the gate function is still effective,
click [Transform] option button in Time Domain Transform dialog box and click to clear [Transform] check box.
8)
Click [OK] to close Time Domain Transform dialog box after setting.
6.9.2. By front panel keys—Time Domain Transform Dialog Box
1)
Press 【Analysis】in MKR/ANALYSIS area, press soft-key corresponding to [Transform], relative soft-key menu appears,
press corresponding soft-key of [Transform ...] to display Time Domain Transform dialog box.
2)
Press 【Tab】 in ADJUST area to switch to [Transform] check box, press 【Click】 to select check box to Stimulus time
domain transformation function.
3)
Press 【Tab】 to switch to Transformation Mode zone, Press 【↑】 , 【↓】 in ADJUST area or rotate knob to select [Low
Pass Impulse], [Low Pass Step] or [Band Pass] mode.
4)
Press 【Tab】to switch to [Start], [Stop] or [Center], [Span] box to set time domain Measure range.
5)
Set window function: press 【Tab】 to switch to [Transform] option button, press 【↑】, 【↓】 or rotate knob to switch
to [Window] option button and then set window filter in three methods below:
Fig.6-27 Time Domain Transformation Measure—Setting Window
Fig.6-28 Time Domain Transformation Measure—Gate Setting
Tel: 886.2.2175 2930 Email: sales@salukitec.com
112
www.salukitec.com