Saluki S3601 Series User Manual page 54

Vector network analyzer
Table of Contents

Advertisement

By front panel keys
4.6.2.2.
1)
Press 【Trigger】 Key in STIMULUS area to show corresponding soft-key menu.
2)
Press the soft-key corresponding to [Trigger...] to show Trigger dialog box.
3)
Complete trigger settings in the dialog box.
4)
After setting, press 【OK】 key in ENTRY area to close the dialog box.
Trigger dialog box
4.6.2.3.
Trigger Source
Trigger source settings determine where the trigger signal is from. Effective trigger signal will yield only when the Analyzer is not
Sweep. The Analyzer may select 3 types of trigger source: internal, manual and external. Once the trigger source is set, the Analyzer
will apply this trigger source to all channels.
[Internal] (default)
 The Analyzer will automatically control the generation of trigger signal.
 The Analyzer will generate a trigger signal immediately after a Measure.
[Manual]
 Manual trigger signal may be created by the following averages:
 Click [Manual Trigger] in Trigger dialog box.
 Press the Trigger key in front panel and press the corresponding soft-key to [Manual Trigger] in soft-key toolbar.
 Call program control command.
[External]
 Trigger signal is input from external trigger input connector on the rear panel.
 The signal is TTL level.
Fig. 4-21 Trigger Dialog Box
54
Tel: 886.2.2175 2930 Email: sales@salukitec.com
www.salukitec.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

S3601aS3601b

Table of Contents