Enhanced Response Calibration; 1- Port (Reflection); Quick Solt - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

Crosstalk
Measure application:
Transmission Measure in any direction.
Measure requiring isolation calibration to improve system dynamic range.
If each port can not connect one load at the same time, isolation calibration can not be conducted.

5.1.5. Enhanced Response Calibration

Calibration precision: medium to high.
Measure parameters: S12 and S21 or S22 and S12.
The required calibration standards: through standard, open-circuit device, short-circuit device and load.
The corrected system errors:
Directivity
Source matching
Reflection Tracing
Transmission Tracing
Measure application:
Transmission Measure in any direction.
Reflection Measure on any port.

5.1.6. 1- Port (Reflection)

Calibration precision: high.
Measure parameters: S11 and S22.
The required calibration standards: open-circuit device, short-circuit device and load.
The corrected system errors:
Directivity
Source matching
Reflection Tracing
Measure application:
Reflection Measure on any port.

5.1.7. Quick SOLT

Calibration precision: medium.
Measure parameters: all.
The required calibration standards: open-circuit device, short-circuit device, load and through standard.
The corrected system errors:
Directivity
Tel: 886.2.2175 2930 Email: sales@salukitec.com
75
www.salukitec.com