 Press 【Tab】 to switch to sliding block box, press 【↑】 , 【↓】 or rotate knob till pulse width or sidelobe level meets
the requirement.
 Press 【Tab】 to switch to [Gaesarian Window β Para] to adjust setting of value β, till pulse width or sidelobe level
meets the requirement.
 Press 【Tab】 to switch to [Pulse Span] to set response pulse width directly.
6)
Set filtering function of time domain gate: press 【Tab】 to switch to [Window] option button, press 【↑】 , 【↓】 or rotate
knob to switch to [Gate Setting] option button, press【Tab】 to switch to [Gating] check to, press 【 Click】 to select the check
box and Stimulus filtering function of time domain gate, then set the function of time domain gate.
 Press 【Tab】to switch to [Start], [Stop] or [Center], [Span] to set gate border.
 Press 【Tab】 to switch to [Gate Type] box, press 【↑】 , 【↓】 or rotate knob to select gate type: band pass, band
stop.
 Press【Tab】 to switch to [Gate Shape] box, press【↑】 ,【↓】 or rotate knob to select gate form: minimum, standard,
wide, maximum.
7)
If it is necessary to observe frequency domain response of the measured piece when the gate function is still effective,
press 【Tab】to switch to [Gate Setting] option button, press 【↑】, 【↓】or rotate knob to switch to [Transform] option
button, press 【Tab】 to switch to [Transform], press 【Click】to clear the check box and close time domain transformation
function.
8)
Press 【OK】in Input keypad to close Time Domain Transform dialog box after setting