Source matching
Reflection Tracing
Crosstalk
Transmission Tracing
Load matching
Measure application:
Measure of all S parameters.

5.1.8. Full Dual-port SOLT

Calibration precision: high.
Measure parameters: all.
The required calibration standards: open-circuit device, short-circuit device, load and through standard.
The corrected system errors:
Directivity
Source matching
Reflection Tracing
Crosstalk
Transmission Tracing
Load matching
Measure application:
Measure of all S parameters.
Measure requiring 12 errors correction to improve the Measure precision.

5.1.9. Full Dual-port TRL

Calibration precision: very high.
Measure parameters: all.
The required calibration standards: through standard, reflection standard, and transmission standard.
The corrected system errors:
Directivity
Source matching
Reflection Tracing
Crosstalk
Transmission Tracing
Load matching
Measure application:
Measure of all S parameters.