5. Calibration

5.1.Selecting Calibration Type

S3601 Series Vector Network Analyzer supports 7 calibration types. The detailed information on each calibration type is as follows:

5.1.1. Open Response

Calibration precision: low to medium.
Measure parameters: S11 and S22.
The required calibration standard: open-circuit device.
The corrected system errors:
Reflection Tracing
Measure application:
Reflection Measure on any port.

5.1.2. Short Response

Calibration precision: low to medium.
Measure parameters: S11 and S22.
The required calibration standard: short-circuit device.
The corrected system errors:
Reflection Tracing
Measure application:
Reflection Measure on any port.

5.1.3. Through Response

Calibration precision: medium.
Measure parameters: S12 and S21.
The required calibration standard: through standard.
The corrected system errors:
Transmission Tracing
Measure application:
Transmission Measure in any direction.

5.1.4. Through Response and Isolation

Calibration precision: medium.
Measure parameters: S12 and S21.
The required calibration standards: through standard, two loads (each port needs one load).
The corrected system errors:
Transmission Tracing
Tel: 886.2.2175 2930 Email: sales@salukitec.com