Y axis unit: time (second).
Suitable typical Measure: group delay.
Linear magnitude format
Only positive value will be displayed.
Y axis: no unit (U) when ratio is measured and the unit is watt (W) when non-ratio power is measured.
Suitable typical Measure: reflectance and transfer factor (magnitude value), time domain transformation.
SWR format
It display reflection Measure data through the formula:(1+ρ)/(1-ρ), in which ρ is the reflectance.
It is effective only reflection Measure is conducted.
Y axis: no unit.
Suitable typical Measure: standing wave ratio format.
Real format
Display the real part of measuring complex data.
Similar with linear magnitude format, but may display positive and negative value.
Y axis: no unit.
Suitable typical Measure: time domain Measure, auxiliary input voltage signal Measure for maintenance purpose.
Imaginary format
Display the imaginary part of measuring complex data only.
Y axis: no unit.
Suitable typical Measure: impedance Measure when matching network is designed.
Polar coordinate format
4.7.1.2.
Polar coordinate format includes magnitude and phase information, and the vector value is read by the following averages:
The magnitude for any point is determined by the displacement from this point to the center point (or zero point). By default, the
magnitude is in linear scale and the outer circle is set to be at the ratio 1.
The phase value of any point is determined by its included angle to X axis.
As there is no frequency information, frequency information is read through marker. By default, the marker format is
real/imaginary part. The other formats may be selected through the Marker dialog box in Marker menu.