Calling Back File - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

(default.cst) of Analyzer, if default file already exists, a dialog box is displayed to ask if user would cover it.
3)
Press corresponding soft-key of [Save As] to display Save As dialog box, and save file according to user setup.
Save As dialog box
7.1.1.3.
[Save In:]
It is used to display and set up file storage path.
[File List]
It is used to display folder under current path and save all type matching files; if click a file among them, file name of save data can
be setup, and if double-click folder, the current path can be modified.
[File Name]
It is used to set up saved file name, display input file name or clicked file name in [Listed File] box.
[Save as Type]
It is used to select file save type, and the Analyzer supports the following file types:
cst file: to save Analyzer status and calibration data.
sta file: only to save Analyzer status data.
cal file: only to save calibration data of Analyzer.
dat file: If user selects to save dat type file, when clicking [Save] button, Analyzer will open Data Save Setting dialog box, it
will save Trace data as per user requirements.
s1p and s2p files: If user selects to save s1p or s2p type files, when clicking [Save] button, Analyzer will open Designated
Data Storage Format dialog box, and it will save Trace data as per user requirements.
prn file: only to save current Stimulus Trace data.
bmp file: to save screen display information of Analyzer in BMP format.
jpg file: to save screen display information of Analyzer in jpeg format.

7.1.2. Calling back file

It can save Analyzer status and calibration data, if necessary it will be called-back in the future. Analyzer supports the following 3
types of files for call-back.
sta file:status data of Analyzer is saved in sta files, including settings, Trace data, limit line and marker of Analyzer.
cal file: calibration data is only saved in cal file, excluding status data of Analyzer. Correction precision of calibration data is
relevant to status setting of Analyzer; therefore, in order to obtain the highest Measure precision, it shall be ensured that status
settings of Analyzer during cal file call-back consistent with that during calibration, otherwise calibration precision can not be
guaranteed.
cst file: cst file saves all Measure status and calibration data of Analyzer, therefore calling cst file can save test time and
improve Measure accuracy.
By mouse or touch screen
7.1.2.1.
1)
Click [File], and click [Recall] on file menu to display Open dialog box.
2)
Select call-back file type in [Files of Type] box.
3)
Set up directory of call-back file through [Look in:] box and the following [File name] box.