Saluki S3601 Series User Manual page 52

Vector network analyzer
Table of Contents

Advertisement

6)
After setting, click [OK] to close the dialog box.
By front panel keys
1)
Press 【Sweep】 key in STIMULUS area to show corresponding soft-key menu. Press the soft-key corresponding to
[Sweep Setup...] to show Sweep Setup dialog box.
2)
Press 【Tab】 in ADJUST area to [Channel]. Press 【↑】 , 【 ↓】 in ADJUST area or turn the knob to select channel setting.
3)
If the default Step Sweep mode is adopted, press 【Tab】 in [Dwell Time] to set dwell time of each measuring point.
4)
If Alternate Sweep mode is to be selected, press 【Tab】 in [Alternate Sweep], and press 【Click】 to select the check
box.
5)
If external fixed amplitude is needed, press【Tab】 in [External ALC], and press【Click】 to select the check box.
6)
After setting, press 【OK】in ENTRY area to close the dialog box.
Sweep Setup dialog box
4.5.3.3.
[Channel] box:
Select the channel adaptable to Sweep Setup.
[Step Sweep] check box
When Step Sweep mode is selected, the source will be tuned to a frequency measuring point, waiting for measuring of this frequency
point after a specified dwell time, and then it will be tuned to the next frequency point. This working mode is slower than Analog
Sweep mode, but can conduct precise Measure for devices with long electrical delay. When this check box is cleared, Analyzer may
operate either in Analog Sweep mode or Step Sweep mode, which depends on the setting of Sweep time and IF bandwidth but
preferentially selects Analog Sweep mode based on Measure settings.
[Dwell Time] input box
It will appoint the dwell time of Analyzer at each point before Measure data is obtained, which is suitable for Step Sweep working
mode only.
Fig. 4-18 Sweep Setup
52