Printing Measure Display; Setting Print Content - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

6)
Press 【 】in ENTRY area to display Data Save Setting dialog box.
7)
Press 【Tab】to shift to format selection zone, press 【↑】 and 【↓】 or rotate knob to select data save format.
8)
Press 【 】to close dialog box, and complete data save at the same time.
prn file
7.1.3.3.
prn file save measured data of Stimulus Trace in the form of row and column, and each row corresponds to a measuring point: the
first column corresponds to measured activation value, the second column corresponds to measured response value. Columns are
separated by comma (,), and format is as following:
Saving prn file
By mouse or touch screen
1)
Click [File], and click [Save As...] on File menu to display Save As dialog box.
2)
Set up file storage type as List File (*.prn) in [Save as Type] box.
3)
Set up file save directory via [Save In:] box and [File List] box.
4)
Set up file name in [File Name] box.
5)
Click [Save] button to save file and close the dialog box.
By front panel keys
1)
Press 【Save】in UTILITY area.
2)
Press corresponding soft-key of [Save As] to display [Save As] dialog box.
3)
Press 【Tab】 in ADJUST area to shift to [Save as Type] box, press 【↑】 and 【↓】 in ADJUST area or rotate knob
to select save file type as list file (*.prn).
4)
Set up file save directory via [Save In:] and [File List] box.
5)
Press 【Tab】to shift to [File Name] box, and set up file name.
6)
Press 【 】in ENTRY area to save file and close the dialog box.

7.2.Printing Measure Display

Analyzer can output or print Measure display content to a specified file via printer. Printer can be local or network printer, types
including parallel port, serial port or USB interface printers. Measure print can be realized once addition of printer to Windows XP
operation system is completed.

7.2.1. Setting Print Content

S11
LIN_SWEEP(Hz), LOG_FORMAT(dB)
3.000000e+005,
-9.232986e+000
7.502250e+008,
-3.219671e-001
1.500150e+009,
-6.892332e+000
2.250075e+009,
-1.146303e+000
3.000000e+009,
-1.245240e+001
119
Tel: 886.2.2175 2930 Email: sales@salukitec.com
Log Mag
www.salukitec.com

Advertisement