Effects when using window wave filter is shown below:
The Analyzer performs window filtering to frequency domain Measure data and has following effects on time domain Measure of the
measured piece:
In impulse transformation mode, it affects pulse width and sidelobe level, changing resolution and dynamic range of Measure.
In low-pass step transformation mode, it can mitigate overshoot and ringing.

8.3.2. Setting Window Value Correctly

When performing time domain Measure, proper window value must be selected according to response type of the measured piece:
Response of the measured piece with identical amplitude: small window is selected to obtain narrow pulse width to improve
resolution of time domain Measure.
Response of the measured piece with different amplitudes: large window is selected to obtain low sidelobe level to improve
dynamic range of time domain Measure.
Fig.8-8 Effect of Window toward Response Signal
Fig.8-9 Correct Window Setting
128