7)
Connect calibration standard on the port, and tick the [Show Standard] dialog box in the top left corner of Measure
Mechanical Standard dialog box. When the button corresponding to the calibration standard to be measured (such as
[OPEN]) is clicked, the Multiple Standards dialog box will be displayed. If the average function should be used to reduce
noise effects during Measure of calibration standard, click to tick the [Average] check box, and set suitable average
factors.
8)
If the [Prompt] dialog box in the top right corner of Measure Mechanical Standard dialog box is ticked, when the button
corresponding to the calibration standard, e.g. [N Male (OPEN)], is clicked in the Multiple Standards dialog box, the
Prompt for Connect Cal standard dialog box will be displayed. Click [OK] button to start Measure after confirming the
connected calibration standard is correct, and the Analyzer automatically ticks the corresponding [Acquire] check box in
the Multiple Standards dialog box after Measure is completed. Click [OK] button to close the Multi-calibration Kit Measure
dialog box after completing all the required calibration standard Measures. Button of the calibration standard whose
Measure is completed in the Measure Mechanical Standard dialog box turns into green.
Figure 5-6 Type Information Dialog Box
Figure 5-7 Multiple Standards Dialog Box