Observing Multi-Trace And Enabling Multichannel; Measure Settings For Pre-Configuration - Saluki S3601 Series User Manual

Vector network analyzer
Table of Contents

Advertisement

3)
Press the soft-key corresponding to [Scale] and set suitable scale in Scale Per division.
4)
Press the soft-key corresponding to [Reference Level] and set reference level in Reference Level.
5)
Press the soft-key corresponding to [Reference Position] and set suitable reference position in Reference Position.
Scale settings dialog box
4.7.3.3.
Scale zone
[Scale]
Set the scale.
[Auto Scale]
Click [Auto Scale] button, the Analyzer will select automatically vertical scale to better show the Active Trace in the vertical grid of the
screen. The stimulus value is not influenced, but the scale and reference value will be changed.
 The Analyzer will select suitable scaleality factor to show the data in 80% zone of the screen.
 If the reference position is at the center of the screen, the reference value will select the center value of the Trace on
the screen.
[Auto All]
Click [All Auto Scale] button, the Analyzer will set suitable scale for all Traces in the window to better show them in the vertical grid of
the window.
Reference setting zone
[Position]
Reference position refers to the position of the reference line in rectangular chart. In the chart, the position of the line at the bottom is
0, the position of the line at the top is 10 and the default value at reference position is 5.
[Level]
Reference value refers to the value of reference line in rectangular coordinates format and the value of outer circle in polar
coordinate and Smith Chart format. Its setting range in log-magnitude format is -500dB~+500dB.

4.8.Observing Multi-Trace and Enabling Multichannel

4.8.1. Measure Settings for Pre-Configuration

The pre-configuration of Analyzer provides a most convenient way to conduct Measure for multi-Trace, multichannel and
multi-window. The Analyzer provides for four Measure settings for pre-configuration, each of which, when selected, shall close the
current Trace and window and create new Trace and window. The arrangement for the window of the four pre-configurations is as
follows: