Fluorescence Detectors FLD-3100 and FLD-3400RS

3.4.3 Connecting the Power Cord

Use the power cord shipped with the device to connect the detector to the main power
source. No manual adjustment is required to adapt the line voltage to local voltage
requirements.
Warning:
Avertissement: Utilisez uniquement les cordons d'alimentation électrique spécifique

3.4.4 Connecting the Analog Outputs (Optional)

3.4.4.1 Installing the DAC Board
1. Turn off the detector and disconnect the power cord from its source.
2. Use appropriate grounding protection to avoid electrostatic discharge, which may
result in damage to electronic components.
3. Loosen the four screws indicated in Fig. 9 and remove the access panel from the rear
panel.
Page 34
Never use a power cord other than the power cords provided for the
device.
Do not use multiple sockets or extension cords. Using defective
multiple sockets or extension cords may cause personal injury or
damage to the device.
à l'instrument.
N'utilisez pas des blocs multiprise ou des câbles prolongateurs. Cela
pourrait entraîner des blessures corporelles ou endommager
l'instrument.
Fig. 9: Screws on the access panel
Operating Instructions