3. Right-click the detector in the timebase and select Properties on the menu.
4. On the General page (→ page 40), the firmware version provided by Chromeleon for
the detector is displayed in the Firmware field. If more than one firmware version is
available for the detector in Chromeleon, select the version from the Firmware list.
5. Click Download. A message displays the firmware version that is currently installed
in the detector and the version that will be downloaded from Chromeleon.
Tip: If the detector comes with a newer firmware than the version included in
Chromeleon, do not downgrade the firmware. Older firmware may be
incompatible with new hardware revisions.
6. Click Yes to start the download. (Click No to cancel the action.)
The download may take several minutes. The download is complete when Download
finished successfully appears in the Messages Server window in the Chromeleon Server
Configuration program. The message appears also in the Chromeleon Audit Trail.
If the download is not successful, the related messages appear in the Audit Trail. In this case,
turn off the detector. Turn on the detector again and repeat the download as described above.
If the download fails again, contact Service.