Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 58

Fluorescence detectors
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
Fluorescence Detectors FLD-3100 and FLD-3400RS
2. Insert the flow cell straight into the cell opening.
3. Tighten the flow cell screws hand-tight.
4. Remove the dummy plugs at the flow cell in and out connection ports. Connect the
flow cell inlet to the separation column. Connect the PEEK capillary from the
accessories kit for the detector as waste capillary to the flow cell outlet. Extend the
PEEK capillary with the silicone tubing (also included in the accessories kit for the
detector). Route the capillaries to the outside through the slots provided in the detector
enclosure to prevent them from being pinched when the front panel door is opened.
5. Close the front panel door to allow the firmware to detect the new flow cell.
6. Perform a wavelength calibration (→ page 90).
Page 50
Fig. 17: Inserting the flow cell
Fig. 18: Tighten the flow cell screws
Flow cell screws
Operating Instructions