4. Slide the left guide bar of the DAC board onto the green edge next to the USB port.
USB port
Guide bar
5. Push the board partially into the enclosure. Apply slight pressure between the two
analog outputs to push the board further into the enclosure until it locks into position.
6. Attach the board to the rear panel, by using the 4 screws as shown in Fig. 9.
7. Connect the analog outputs to an evaluation device (→ page 35).
Tip:
Installation of the DAC board does not require additional validation, such as
Operational Qualification or Performance Qualification.
3.4.4.2 Connecting an Analog Output to an Evaluation Device
1. Locate the analog output cable (part no. 6074.0002) provided with the DAC board.
Plug the cable connector into the analog output on the detector rear panel.
2. Connect the free end of the cable to the analog input on the evaluation device.
3. Make the required settings in Chromeleon (→ section 3.4.4.3).
Operating Instructions
Fluorescence Detectors FLD-3100 and FLD-3400RS
Fig. 10: Installing the DAC board