Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 49

Fluorescence detectors
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

♦ Click Write to save the data currently delivered by the detector as a demo file. Enter
the file name in the Virtual Mode File Name field or select a name from the list.
Tip: When recording virtual mode files in Chromeleon 6.80, the size of the file
is limited to 35MB. This corresponds to about 9 minutes at a data
collection rate of 100 Hz. If you want to record longer virtual mode files,
reduce the data collection rate.
Module Address
Select the module address of the detector. Click Browse and then double-click the
detector that you want to use on the Device List. The address is automatically entered in
the Module Address field. Chromeleon connects to the detector and transfers the settings
from the instrument firmware to Chromeleon, setting the options on the pages
accordingly. Confirm the related message with OK.
Download
Click this button to transfer the current detector configuration to Chromeleon. (The
button appears dimmed if the virtual mode is enabled.)
The detector is shipped with the most recent firmware version. If a firmware update is
ever required, follow the steps in section 7.6 (→ page 115).
Operating Instructions
Fluorescence Detectors FLD-3100 and FLD-3400RS
UltiMate 3000 Series:
Page 41