Configuring The Detector - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Fluorescence detectors
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
Fluorescence Detectors FLD-3100 and FLD-3400RS

3.5.3 Configuring the Detector

3.5.3.1 Initial Installation
During the installation, Chromeleon connects to the detector and transfers the settings from
the instrument firmware to Chromeleon, setting the options on the wizard pages
accordingly. Verify that the settings are correct and make additional settings if needed.
You may reopen the configuration pages later again to change the settings (→ page 43).
Tip: Changing the settings for a specific application in the Commands dialog box,
in a program file (PGM), or on a control panel will not change the default
settings on the configuration pages.
For additional information about a page, click Help.
General Page
Shows the general instrument parameters.
Virtual Mode
Verify that the virtual mode is set to off. If the virtual mode is enabled, the Module
Address box will be unavailable. If you exit this page without having entered a module
address, the virtual mode will be enabled automatically.
In the virtual mode, Chromeleon simulates detector control and data acquisition.
♦ Click Read to read and display data from an existing demo file instead of real data.
Select the file from the Virtual Mode File Name list.
Page 40
Fig. 11: General page
Operating Instructions