OPERATION
6
Menu
Function
No.
2.0.0 Supervisor
2.1.0 Identification
2.1.0 Identification
2.1.0 Identification
2.1.0 Identification
2.1.1
Serial Number
2.1.2
Conv Firm
Version
2.1.3
Sens Firm
Version
2.1.4
HMI Firm Version
2.1.5
Electronic
Revision
2.1.6
Mechanical
Traceability
2.2.0 Tests
2.2.0 Tests
2.2.0 Tests
2.2.0 Tests
2.2.1
Set Curr.Output
I1
2.2.2
Diagnostic
2.2.3
Set Current
Output I2
2.2.4
Switch Output
test
2.2.5
Proof Test
2.3.0 Basic Parameters
2.3.0 Basic Parameters
2.3.0 Basic Parameters
2.3.0 Basic Parameters
2.3.1
Tank Height
90
Function description
The device serial number.
The converter firmware version.
The sensor firmware version.
The HMI (device display screen) firmware
version.
Electronic Revision is a series of numbers
Electronic Revision
Electronic Revision
Electronic Revision
used to record the revision status of
embedded software (firmware) in
electronic equipment assemblies. It gives
data on the type of changes made and the
effect that changes have on compatibility.
The serial number for the probe and
process connection.
Set Current Output. This sets analogue
output 1 to a test value [mA] selected from
a list. Output will change to the selected
value, independent of the measured value.
This starts the hardware test. Push [ ]
and [ ] many times to show: the
functioning time, converter temperature,
current on loop, load current , voltage
5.3 V, voltage on capacitors, voltage 3.3 V,
reference pulse amplitude, flange pulse
amplitude, level pulse amplitude, probe
end pulse amplitude, reset counter and
device status (this includes warning and
error messages that are "on" at this time).
Set Current Output. This sets analogue
output 2 to a test value [mA] selected from
a list. Output will change to the selected
value, independent of the measured value.
This function sends a test signal to show
that the relay is open or closed.
Not yet available. If the device has the SIL
option, it is necessary to do proof tests to
make sure that the safety function is
applicable to the full measuring range. We
recommend that you do a proof test
immediately after you install and start the
device.
The distance from the tank connecting
flange face / thread stop down to the tank
bottom.
www.krohne.com
OPTIFLEX 8200 C/F/S
Selection list or range
of values
Read only.
Read only.
Read only.
Read only.
Read only.
Read only.
3.5, 4, 6, 8, 10, 12, 14, 16,
18, 20 or 22 mA
Read only.
3.5, 4, 6, 8, 10, 12, 14, 16,
18, 20 or 22 mA
OPEN, CLOSE
min-max:
0...99.999 m /
0...328.08 ft
06/2018 - 4006448901 - MA OPTIFLEX 8200 R01 en
Default
3.5 mA
3.5 mA
OPEN
If tank height is
not specified in
customer order,
the probe
length value is
used