ELECTRICAL CONNECTIONS
4
How to connect the signal cable to the remote converter
Figure 4-17: How to connect the signal cable to the remote converter
CAUTION!
Bending radius of the signal cable: ≥ 50 mm / 2
1 Remove the terminal compartment cover.
2 Remove the 4-pin connector.
3 Put the signal cable into the opening of the cable gland.
4 Put the electrical wires in the connector terminals. Tighten the terminal screws with a small
slotted-tip screwdriver. Make sure that the electrical wires agree with the terminals. For
more data, refer to the electrical schematic diagram in this section.
5 Put the connector into the 4-pin socket.
6 Attach the Faston connector (drain wire).
7 Attach the terminal compartment cover.
8 Tighten the cable gland. Make sure that the remote converter is correctly sealed.