Part 2: How to assemble the segmented single rod probe
Figure 3-18: Part 2: How to assemble the segmented single rod probe
WARNING!
Put a support below the probe to prevent deformation.
CAUTION!
Make sure that the nuts are tight and the rod probe cannot loosen.
1 Attach a lock nut to each end of the rod probe segments.
2 Attach a union nut to the bottom end of each rod probe segment, but not to the bottom seg-
ment. Use two 8 mm open-end wrenches to tighten the union nut against the lock nut.
3 Attach the top segment of the rod probe to the union nut below the process connection. Use
two 8 mm open-end wrenches to tighten the union nut against the lock nut on the rod probe.
4 Attach the middle segment of the rod probe to the union nut on the top segment (if there are
middle segments). Use two 8 mm open-end wrenches to tighten the union nut against the lock
nut. Repeat this step for the other segments.
5 Attach the bottom segment of the rod probe to the union nut on the top segment. Use two
8 mm open-end wrenches to tighten the union nut against the lock nut.