5.2 Before and after opening....................29 5.3 Replacement of mains fuse.................... 30 5.4 Exchange of electronics unit ..................31 5.4.1 Field version.......................... 32 5.5 Service / repair information ................... 34 www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 3
CONTENTS OPTISONIC 7300 / 8300 5.6 Form (for copying) to accompany a returned device ............. 35 5.7 Disposal .......................... 35 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence. RESULT RESULT RESULT RESULT This symbol refers to all important consequences of the previous actions. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Explosion protected versions of the devices. All technical information as described in the standard documentation is applicable, when not specifically excluded, completed or replaced by the instructions in these additional instructions. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
DEKRA 12 ATEX 0063 X 1.4 OPTISONIC 7300 C/...-Ex The OPTISONIC 7300 C/...-Ex is the compact configuration of the GFC 300 ultrasonic flow converter and the OPTISONIC 7000 ultrasonic gas flow sensor. The gas flow sensor contains flameproof (Ex d) gas transducers made of titanium or flameproof (Ex d) transducers with a partly encapsulated (Ex ma) synthetic front.
These transducers have marking code: II 1/2 G Ex d ma IIC T4...T6. The high temperature transducers (OPTISONIC 8000-Ex) are always flanged and have type of protection flameproof enclosure (Ex d). See Chapter 2 for details on the temperature classification table(s). 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
(this is the temperature of the medium that surrounds the transducers). The ambient temperature is the (air) temperature at the location where the meter is installed. 1.10 Marking codes Due to the various configuration options, the OPTISONIC 7300 gas flowmeter series can have several (different) explosion safety measures implemented. www.krohne.com...
Page 11
II 2 G Ex d [ib] IIC T6 Ex d n.a. Ex ib II 2 G Ex de [ib] IIC T6 Ex de n.a. Ex ib 1 Only available in separate version. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
3 Tag number. 4 PED data. 5 Calibration data. 6 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies. 7 Name and address of the manufacturer. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 13
5 Main supply data. 6 Calibration data. 7 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies. 8 Name and address of the manufacturer. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
Page 14
5 Main supply data. 6 Calibration data. 7 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies. 8 Name and address of the manufacturer. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 15
5 Main supply data. 6 Calibration data. 7 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies. 8 Name and address of the manufacturer. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
• die-casted stainless steel: -40...+55ºC 2.3 OPTISONIC 7300 C/...-Ex The OPTISONIC 7300 C/...-Ex ultrasonic compact gas flow meter is provided with the same electronics housing as the GFC 300 F/...-Ex and therefore has the same ambient temperature range of -40...+65°C (aluminum housing) and -40...55°C (stainless steel). The temperature classification table shows the relationship between the maximum process medium temperatures at a certain maximum ambient temperature.
The OPTISONIC 8000...-Ex high temperature (steam) flow sensor has an ambient temperature in the range of -40...+60°C. The maximum process medium temperature is listed in the table below. Temperature class Maximum process temperature [ºC] at T = 60ºC 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
(1.1, 1.2, ...). 3.2.1 Sensor Figure 3-1: Connect the cables on the sensor. 1 Sensor. 2 Open connection box. 3 Marking on cable. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 19
Figure 3-2: Clamp the cables on the shielding bush. 1 Cables. 2 Cable glands. 3 Grounding clamps. 4 Cable with metal shielding bush. Figure 3-3: Connect the cables on the signal converter. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
3 Marking on cable. 4 Insert cables through cable glands. Figure 3-4: Clamp the cables on the shielding bush. 1 Cables. 2 Cable glands. 3 Grounding clamps. 4 Cable with metal shielding bush. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
In the special case of a screen being earthed at both ends of the system, a potential difference between the sensor and converter is not permitted. Refer to EN 60079-14, Clause 12.2.2.3, for further requirements of cable screens. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
WARNING! Allow the electronics to de-energize before opening the electronics compartment of the flow converter housing. Wait at least 35 minutes for T6 and 10 minutes for T5 before opening. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 23
Connections for signal I/Os (PELV circuits), non-“Ex i” or “Ex i”, B, B- are dependent on the specific version of the GFC 300 C, C- converter ordered. Consult the tables with CG36 numbers for D, D- details. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
Page 24
11 (additional requirements for type of protection “e” - increased safety) and section 12 (additional requirements for type of protection “i” - intrinsic safety) apply to respectively “Ex d”, “Ex e” and “Ex i” performed connection (terminal) compartments. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Carrier with 1 or 2 I/O modules Fieldbus I/O n.c. n.c. Foundation Fieldbus E8A to ELL Fieldbus I/O Many combinations possible Profibus PA with Module Carrier with 1 or 2 I/O modules 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
Page 26
CI = Control Input, PA = Profibus PA, FF = Foundation Fieldbus, DP = Profibus DP, RS485 = RS 485 Modbus, n.c. = not connected. All in-/outputs are passive unless otherwise noted as active with extension (a). www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
60079-11 between the non-”Ex i” and “Ex i” circuits, the mains terminals are provided with a semi-circular protection cover with a “snap-in” lock. This cover MUST be closed before establishing the power supply to the converter. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
Page 28
Used abbreviations for in-/output functions: CO = Current Output, PO = Pulse Output, SO = Status Output, CI = Control Input, n.c. = not connected All in-outputs are passive unless otherwise noted as active with extension (a). www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
• Screw the cover as tight as possible onto the housing by hand, until it cannot be opened by hand anymore. Fixate the screw of the interlocking device tight with the No. 3 Allen key. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
100...230 V AC 250 V / 0,8 A WARNING! Before reassembling the unit, refer to Before and after opening on page 29 , then: • Reassemble the unit in reverse order. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Only the sensor calibration data are loaded. - if in the screen appears “load no data load no data”, all data have been lost. Contact your local load no data load no data representative. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
CAUTION! Please pay attention that the same amount of force is applied on both pullers, otherwise the connector at the backside can be damaged. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
Page 33
• Screw the electronics unit back to the housing. • Re-install the display and make sure not to kink the display's flat ribbon cable. • Replace cover and tighten by hand. • Connect power. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
• such dangerous substances, to enclose a certificate with the device confirming that is safe to handle and stating the • product used. www.krohne.com 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02...
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned. Date: Signature: Stamp: 5.7 Disposal CAUTION! Disposal must be carried out in accordance with legislation applicable in your country. 05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02 www.krohne.com...
• Measuring systems for the marine industry Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de The current list of all KROHNE contacts and addresses can be found at: www.krohne.com...