INSTALLATION
3
Part 3: How to attach the locking nut and union nut
Figure 3-15: Part 3: How to attach the locking nut and union nut
CAUTION!
Support the probe.
1 Attach a locking nut to the single rod.
2 Make sure that the locking nut is engaged ¾ along the length of the thread.
3 Attach the single rod to the union nut. Make sure the probe touches the threaded rod below
the process connection.
4 Tighten the single probe with a 7 mm open-ended wrench.
5 Tighten the locking nut against the union nut with two 8 mm open-ended wrenches.
34
www.krohne.com
OPTIFLEX 8200 C/F/S
06/2018 - 4006448901 - MA OPTIFLEX 8200 R01 en