Pulse Sources - Agilent Technologies E8241A User Manual

Psg series signal generators
Hide thumbs Also See for E8241A:
Table of Contents

Advertisement

Fundamental Operation
Configuring Analog Modulation (PSG-A Series Only)

Pulse Sources

The following list summarizes the different sources available for pulse modulation.
External Pulse
Internal Doublet
Internal Free-Run
Internal Gated
Internal Trigger
Internal Square
For a detailed description of the behavior, ranges, and limitations of a specific pulse source,
see that waveform's softkey description in
34
modulates by following an external pulse signal connected
to the signal generator's PULSE/TRIGGER GATE INPUT
connector.
produces two pulses at the RF OUTPUT connector for
each trigger event at the TRIGGER IN connector. The first
pulse will follow the external trigger signal. The second
pulse will have user-defined delay and width parameters.
produces internal, free-run, pulse modulation with
user-defined period, width, and delay.
produces an internal, gated, pulse modulation. When a
valid gate signal is applied to the TRIGGER IN connector,
a pulse train (with user-defined width parameters) will
occur at the RF OUTPUT connector.
produces an internal, triggered, pulse modulation. An RF
pulse (with user-defined width and delay parameters) will
occur at the RF OUTPUT connector whenever a valid
trigger signal occurs at the TRIGGER IN connector.
produces internal, square, pulse modulation. This is a
form of internal free-run pulse with a 50% duty cycle. The
period is determined by rate.
Chapter 5, "Key Reference," on page
91.
Chapter 2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Us4124E8251aE8244aE8254a

Table of Contents