Page 2
Notice The material contained in this document is provided “as is”, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and to any of the Agilent products to which it pertains, including but not limited to the implied warranties of merchantability and fitness for a particular purpose.
Troubleshooting AT1 Attenuator ......... . 1- 45 Agilent E8663B Analog Signal Generator Service Guide...
Page 4
Save Money with Rebuilt- Exchange Assemblies ........3- 1 Shipping a Defective Assembly Back to Agilent Technologies ......3- 2 Frequency, Hardware, and Personality Options.
Page 5
Replacement Procedure ..........2- 10 Agilent E8663B Analog Signal Generator Service Guide...
Page 6
..........2- 30 Agilent E8663B Analog Signal Generator Service Guide...
Page 7
Removal Procedure........... 2- 54 Agilent E8663B Analog Signal Generator Service Guide...
Page 8
Certification ............5- 2 viii Agilent E8663B Analog Signal Generator Service Guide...
Page 9
Basic Function Commands • Analog Commands • SCPI Command Compatibility Service Guide • Troubleshooting • Replaceable Parts • Assembly Replacement • Post- Repair Procedures • Safety and Regulatory Information Key Help • Key function description Agilent E8663B Analog Signal Generator Service Guide...
Page 10
Agilent E8663B Analog Signal Generator Service Guide...
Always handle a printed circuit board assembly by its edges. This reduces the possibility of ESD damage to components, and prevents contamination of exposed plating. Agilent E8663B Analog Signal Generator Service Guide...
• If the Self Test passed: “The current status of self test is: Success All self tests have passed.” • If the Self- Test failed: “The current status of self test is: Failed.” The message also reports the most independent failure. Agilent E8663B Analog Signal Generator Service Guide...
5. If either of the voltages are missing, refer to “Power Supply Troubleshooting” on page 1- 6. If DS9 and DS10 are on, proceed to “Checking the A18 CPU Voltages” on page 1- Agilent E8663B Analog Signal Generator Service Guide...
Error Messages If error messages are reported after the turn–on check is completed, refer to the Error Messages list, located on the E8663B Signal Generator Documentation Set CD (part number E8663- 90010) for details. Agilent E8663B Analog Signal Generator Service Guide...
1100 Power Supply A26 Micro Interface Deck “Self–Tests 12xx: A26 MID Self–Test Errors” on page 1- 29 1200 Power Supply 1201 Mod Filter 1202 20 GHz Doubler 1204 RF Path 1205 LB Pulse Mod Agilent E8663B Analog Signal Generator Service Guide...
Page 16
“Self–Tests 22xx: A8 Output Self–Test Errors” on page 1- 34 2200 Ground and PTAT Test 2201 Prelevel Loop Test 2202 Switched Filters Test 2203 ALC Mod System Test 2204 Pulse Mod System Test 2205 RF Path Test Agilent E8663B Analog Signal Generator Service Guide...
The A2 Display brightness can be adjusted from 50 (high brightness) to 1 (low brightness) using the RPG, arrow keys, or the numeric keypad. Other display features located under the Utility > Display softkey, include a screen saver and inverse video. Agilent E8663B Analog Signal Generator Service Guide...
Page 18
34, 36, 37, 39, 40 If >21V check the power supplies. If the power supplies are good, replace the motherboard. The pulse state signals are control lines, clock, and data for the LCD. Agilent E8663B Analog Signal Generator Service Guide...
Page 19
J9−1, 2, 3, 8 to 15 Refer Figure 1- 1 on page 1- 9. If these signals are good, replace the LCD. If these signals are not good, replace the power switch/flat panel interface board. Agilent E8663B Analog Signal Generator Service Guide...
Page 20
5. If both signals are correct, press the faulty hardkey or softkey while monitoring the KEYCOL or KEYROW on the oscilloscope. The KEYCOL line should pulse low; the KEYROW line should pulse high. If either line does not function properly, replace the front panel board. 1-10 Agilent E8663B Analog Signal Generator Service Guide...
Page 21
P112- 17, 18, P112- 3, 28 A10 ALC P122- 4 P122- 28 P122- 2, 27 A11 Mod Gen P132- 4 P132- 28 P132- 2, 27 A12 Pulse Modulator J14- 10 J14- 9 1-11 Agilent E8663B Analog Signal Generator Service Guide...
Page 22
–5.2 Vdc –3 Vdc +3.5 Vdc +6.5 Vdc +12 Vdc Voltage A2 Display P11- 43 A3 Power Supply P231- 12, P231- 80 13, 52, 53 A5 Sampler P22- 2, 17 P22- 4, 19 1-12 Agilent E8663B Analog Signal Generator Service Guide...
Page 23
A29 20 GHz Doubler J32- 4 J32- 8 J32- 1, 3 A30 Mod Filter J31- 4 J31- 8 J31- 1, 3 B1 Fan Fan+P25 2- 2 252- 2 Fan- P252- 1 P252- 1 1-13 Agilent E8663B Analog Signal Generator Service Guide...
Page 24
+15 Standby LED on the motherboard. This is controlled by the +15 Standby line supplied by the power supply, which is routed via the motherboard to the front panel. To troubleshoot, turn the instrument off and follow the signal path. 1-14 Agilent E8663B Analog Signal Generator Service Guide...
Page 25
3. Turn the instrument off and reinstall the assembly removed in Step 1. Remove the next assembly and see if the supply LED lights. Continue this process until the supply functions properly. 4. Replace the last assembly that you removed. 1-15 Agilent E8663B Analog Signal Generator Service Guide...
Page 26
1. Turn the front panel switch on and ensure that the +15 Standby is good. 2. Verify that the ON/OFF control line is > 3.5 Vdc. The Power Switch/Flat- Panel Interface Board in the front panel assembly controls this signal. 1-16 Agilent E8663B Analog Signal Generator Service Guide...
Page 27
4. Power the instrument up again and check the faulty supply. If it is still bad, power down and re- install the first assembly removed. 5. Remove the next assembly and see if that fixes the problem. Continue this process until the supply functions properly. 6. Replace the last assembly that you removed. 1-17 Agilent E8663B Analog Signal Generator Service Guide...
3. If the supply voltages are good at the origin, replace the A31 Motherboard. 4. If the supply voltages are bad at the origin, troubleshoot the problem using that assembly’s troubleshooting procedure. 1-18 Agilent E8663B Analog Signal Generator Service Guide...
Page 29
8 GHz CW, the A6 Frac–N signal should be at a frequency of 593.75 MHz and at a power level of +7 dBm. 4. If the signal is good, replace the A5 Sampler, if the signal is bad go to the A6 Frac–N troubleshooting procedure. 1-19 Agilent E8663B Analog Signal Generator Service Guide...
–6.86 –7.14 Main Supply +5.2 Digital high P111- 64, 65, 129, 130 +5.04 +5.36 Main Supply +3.4 Digital Low P111- 60, 61, 62, 63, 125, 126, 127, 128 +3.29 +3.5 Main Supply 1-20 Agilent E8663B Analog Signal Generator Service Guide...
Page 31
Sweep: Freq 1 MΩ Sweep Time: Auto Sweep Repeat: Continuous A positive- and negative- going signal should be present. • If the signal is present at J9, replace the A9 YIG Driver. 1-21 Agilent E8663B Analog Signal Generator Service Guide...
YIG Driver –5.2 P32- 4,19 –5.1 –5.3 YIG Driver –6 P32- 3, 18 –5.88 –6.12 YIG Driver 700 Ref Voltage • Replace the A6 Frac–N. 701 Loop Gain • Replace the A6 Frac–N. 1-22 Agilent E8663B Analog Signal Generator Service Guide...
Page 33
• If the signal out the A29 20 GHz Doubler is good, replace the cable. • If the signal out the A29 20 GHz Doubler is bad, replace the A29 20 GHz Doubler. 1-23 Agilent E8663B Analog Signal Generator Service Guide...
Page 34
• If the signals is present, replace the A6 Frac–N. • If the signal is not present, refer to A11 Pulse/Analog Modulation Generator troubleshooting. 707 Sweep Phase Detector Test 1. Replace the A6 Frac–N. 1-24 Agilent E8663B Analog Signal Generator Service Guide...
P121- 64, 65, 129, 130 +5.04 +5.36 Main Supply +3.4 Digital Low P121- 60, 61, 62, 63, 125, 126, 127, 128 +3.29 +3.5 Main Supply 900 Power Supply • Replace the 10 ALC assembly. 1-25 Agilent E8663B Analog Signal Generator Service Guide...
Page 36
• If the signal is not present, refer to highband RF loop troubleshooting. • If the signal is present, replace the A10 ALC. 904 Scalar Mod Pulse • Replace the A10 ALC. 905 ALC Mod Test • Replace the A10 ALC. 1-26 Agilent E8663B Analog Signal Generator Service Guide...
. If the signal is present, replace the A7 Reference. 2. If the signal is not present, replace the internal 10 MHz standard. 1002 1 GHz Test • Replace the A7 Reference. 1-27 Agilent E8663B Analog Signal Generator Service Guide...
+9 ±.18 Vdc 4. If the voltages are present on the A31 Motherboard, turn power off, remove the A18 CPU and inspect the connector. If the pins look good, replace the A18 CPU. 1-28 Agilent E8663B Analog Signal Generator Service Guide...
P201- 15, 16, 17, 18, 19, 20 +9.7 +10.3 Main Supply P201- 39 +7.5 +8.5 –7 P201- 40, 43, 56 –7.21 –6.79 Main Supply –5.2 P201- 41, 42 –5.30 –5.10 YIG Driver 1-29 Agilent E8663B Analog Signal Generator Service Guide...
Page 40
A24 Highband Coupler and A25 Highband Detector. 1. Connect a spectrum analyzer to the RF output of the signal generator. 2. Turn the RF power on and verify the presence of an RF signal on the output. 1-30 Agilent E8663B Analog Signal Generator Service Guide...
Page 41
Change the pulse period to 1 ms and the pulse width to 500 µs. • If the pulse is not present, perform the A11 Pulse/Analog Modulation Generator self–test, and follow the troubleshooting procedure. 1-31 Agilent E8663B Analog Signal Generator Service Guide...
P131- 64, 65, 129, 130 +5.04 +5.36 Main Supply +3.4 Digital Low P131- 60, 61, 62, 63, +3.29 +3.5 Main Supply 125, 126, 127, 128 1300 Power Supply • Replace the A7 Reference. 1-32 Agilent E8663B Analog Signal Generator Service Guide...
Page 43
• Replace A11 Pulse/Analog Modulation Generator. 1314 External 1 • Replace A11 Pulse/Analog Modulation Generator. 1315 External 2 • Replace A11 Pulse/Analog Modulation Generator. 1316 AM Path Loss • Replace A11 Pulse/Analog Modulation Generator. 1-33 Agilent E8663B Analog Signal Generator Service Guide...
3. Connect a spectrum analyzer to the A6 Frac–N RF output at J4 of the A8 Output. The signal at J4 should be at 3 GHz and ≥0 dBm. • If the signal is not present or at the correct level, replace the Frac–N. • If the signal is present, replace the A8 Output. 1-34 Agilent E8663B Analog Signal Generator Service Guide...
30 minutes. This is not a failure unless the message does not go off after 30 minutes. If it continues to be displayed after 30 minutes, replace the A7 Reference. 1-35 Agilent E8663B Analog Signal Generator Service Guide...
A11 Pulse/Analog Modulation Generator. This error message is turned on if the signal generator is in phase modulation mode and there is a problem with the 10 MHz digital signal to the in phase clock. 1-36 Agilent E8663B Analog Signal Generator Service Guide...
Connect the RF output of the signal generator to the spectrum analyzer. Measure and record the minimum power level. c. Set the signal generator and spectrum analyzer start/stop frequencies to the next start/stop frequency in Table 1- 1-37 Agilent E8663B Analog Signal Generator Service Guide...
Page 48
• Set the signal generator to a frequency between 250 kHz and 3.2 GHz. Set the amplitude so the power meter reads +21 dBm (or the maximum settable power) on cable W27 going to J3 on A30 Modulation Filter. 1-38 Agilent E8663B Analog Signal Generator Service Guide...
Page 49
If the signal generator is unleveled at all frequencies, the problem is most likely the A10 ALC. a. Checking the A10 ALC: • Run a complete self–test and troubleshoot the reported failure. 1-39 Agilent E8663B Analog Signal Generator Service Guide...
ALC Modulator Calibration A30 Modulation Filter A8 Output Attenuator Calibration- Lowband and Highband Power RF Path Attenuator Calibration- Highband and High Power RF Path Attenuator Calibration- Lowband and Low Power RF Path 1-40 Agilent E8663B Analog Signal Generator Service Guide...
6. If self–test fails, turn power off, reinstall the assemblies and remove the ribbon cable from J10 on the A26 MID (W35). 7. If self–test passes, replace the A26 MID. 8. If self–test fails, replace A18 CPU. 1-41 Agilent E8663B Analog Signal Generator Service Guide...
“Option 509 Models” on page 1- Sub- Harmonic Spurious Non- Harmonic Spurious Refer to “Troubleshooting Non–Harmonic Spurious” on page 1- Single- Sideband Phase Noise Refer to “Troubleshooting the RF Path” on page 1- 1-42 Agilent E8663B Analog Signal Generator Service Guide...
10. If signal levels are good, replace the A8 Output. 11. If the signals are bad, check the cable continuity. If the cable is bad, replace the cable. If the cable is good replace the A11 Pulse/Analog Modulation Generator. 1-44 Agilent E8663B Analog Signal Generator Service Guide...
–45.0 to –54.99 dBm –55.0 to –64.99 dBm –65.0 to –74.99 dBm –75.0 to –84.99 dBm –85.0 to –94.99 dBm –95.0 to –104.99 dBm >–105.0 dBm 4. If the voltages are correct, replace the attenuator. 1-45 Agilent E8663B Analog Signal Generator Service Guide...
Use this procedure to troubleshoot harmonic problems with a fundamental frequency between 3.2 GHz and 10 GHz. 1. Set the signal generator as follows: • Frequency: set to harmonic frequency to be measured 1-46 Agilent E8663B Analog Signal Generator Service Guide...
These spurs are generated on the A5 Sampler. They are caused by harmonics of Spurs the A5 Sampler IF. The spurs are measured in highband at 133 kHz offset from the CW frequency. 1-47 Agilent E8663B Analog Signal Generator Service Guide...
This regulation can take place on any assembly and then routed to the required assemblies. 1-48 Agilent E8663B Analog Signal Generator Service Guide...
The rear panel contains the power line module, LAN, GPIB, RS–232, scalar interface, and BNC connectors used for sweep interface. BNC and SMB connectors may be present on the rear panel but not operational if the option associated with that connector is not installed. 1-49 Agilent E8663B Analog Signal Generator Service Guide...
A6 Frac–N frequency accuracy. The frequency reference for the A7 Reference can be an: • external 10 MHz signal • internal high stability 10 MHz OCXO (Oven Controlled Crystal Oscillator) and the A7 Reference 1-50 Agilent E8663B Analog Signal Generator Service Guide...
Page 61
Internally generated signals up to 1 MHz can be switched to the LF Output BNC on the front panel. A11 Pulse/Analog output signals are routed as follows: Modulation Type Routed to Amplitude A10 ALC Pulse A30 Modulation Filter A8 Output Frequency and Phase A6 Frac–N 1-51 Agilent E8663B Analog Signal Generator Service Guide...
Page 62
Troubleshooting Overall Description This page intentionally left blank. 1-52 Agilent E8663B Analog Signal Generator Service Guide...
Page 63
Source Module Crystal Oscillator Interface GPIB Auxiliary Interface Source Settled Trigger In Trigger Out Sweep Out p-anaphoto_top A18 CPU 120 Vac A19 Power Supply A22 Line Module 240 Vac A31 Motherboard Block Diagram for the Overall E8663B Analog Signal Generator e8663b_ovr-all...
FM coils. The main coil is used to set the frequency and for FM ≤230 Hz. The FM coil is used for FM rates >230 Hz. Both the A9 YIG Driver and the A28 YIG Oscillator contribute to phase noise between 100 KHz and 1 MHz. 1-56 Agilent E8663B Analog Signal Generator Service Guide...
≤3.2 GHz; and the A5 Sampler to provide the feedback needed for fine tune correction in CW mode • RF signal amplification and filtering Both the A29 20 GHz Doubler and the A30 Modulation Filter contribute to phase noise >3.2GHz. 1-57 Agilent E8663B Analog Signal Generator Service Guide...
Page 67
Troubleshooting Reference/Synthesis Loop Description This page intentionally left blank. 1-58 Agilent E8663B Analog Signal Generator Service Guide...
Page 68
1 GHz 618 to 905 MHz Output to 900 MHz A5 Sampler 155 MHz Vtune Mixer IF 12 to 155 MHz 750 MHz 900 MHz 750 MHz B Block Diagram for the E8663B Analog Signal Generator Reference / Synthesis Loop E8663b_ref_synth-loop...
UNW hardware, and the pulse diode on the A8 Output is biased on, allowing the RF signal to pass through. The pulse shaper circuitry on the A11 Pulse Modulator needs to be set using a license key to provide narrower pulse widths. 1-61 Agilent E8663B Analog Signal Generator Service Guide...
In external leveling operation, the internal detectors are replaced with an external detector, so the level control voltage is provided by the external detector. The external detector must be a negative detector to provide the correct 1-62 Agilent E8663B Analog Signal Generator Service Guide...
ALC diode drive level. Because drift in that leveling circuit can occur over time in open- loop operation, it is not recommended to stay in search mode for extended periods of time without closing the loop and refreshing the ALC diode drive voltage. 1-63 Agilent E8663B Analog Signal Generator Service Guide...
J1 to J4 Gain > 9 J3 to J4 Loss < 3 A24 Highband Coupler < 3.2 GHz Loss < 2.5 2–9 GHz < 0.5 AT1 (0 dB step) Loss < 2.5 1-64 Agilent E8663B Analog Signal Generator Service Guide...
Page 73
Divide Detector H_BURST Front Panel ALC_MOD_OFFSET_DAC 50 ohm from from 3.2-10GHz ALC_MOD_DRIVER_ A11 Pulse / Analog A7 Reference from MOD_MUX BIAS/GAIN_DAC Modulation Generator A29 Doubler from A10 ALC and A5 Sampler Block Diagram for the E8663B Analog RF Path E8663b_rf-path...
Select/Deselect Test • Run Highlighted Tests • Select/Deselect All • Run Selected Tests • View Details • When View Details is selected, the highlighted test’s measured value, lower, and upper limits are displayed. 1-67 Agilent E8663B Analog Signal Generator Service Guide...
Describe the problem: What action has been taken: If you are having a problem with performance tests or adjustments Software revision: Name of adjustment or performance test: Reported software message: Describe the problem: 1-68 Agilent E8663B Analog Signal Generator Service Guide...
Page 76
Troubleshooting Contacting Agilent Technologies If you are having a problem with an application Describe what you want to do: Describe the problem: 1-69 Agilent E8663B Analog Signal Generator Service Guide...
3. Seal the shipping container securely with strong, nylon adhesive tape. 4. Mark the shipping container “FRAGILE, HANDLE WITH CARE” to ensure careful handling. 5. Retain copies of all shipping papers. 1-70 Agilent E8663B Analog Signal Generator Service Guide...
1. Determine the part number. 2. Determine the quantity required. 3. Mail this information to the nearest Agilent Technologies office or, in the U.S., call the hot- line number listed in the following section. To order a part not listed in the replaceable parts lists, mail the following information to the nearest Agilent Technologies office or, in the United States, call the hot- line number listed in the following section.
If you are inside the United States, stick the pre- printed return address label over the label that is already on the box and return the box to Agilent Technologies. (Agilent Technologies pays postage on boxes mailed within the United States.) If you are outside the United States, do not use the return address label;...
Replaceable Parts Frequency, Hardware, and Personality Options Frequency, Hardware, and Personality Options This section lists the options available with the Agilent E8663B analog signal generator. Frequency Options 100 kHz to 3.2 GHz frequency range 100 kHz to 9.0 GHz frequency range...
Replaceable Parts Frequency, Hardware, and Personality Options Assemblies from the Inside Front Panel View Figure 2-2 Assemblies from the Inside Front Panel View Agilent E8663B Analog Signal Generator Service Guide...
Replaceable Parts Assemblies and Hardware Options Assemblies from a Top View Figure 2-3 Assemblies from a Top View A25B Agilent E8663B Analog Signal Generator Service Guide...
Replaceable Parts Frequency, Hardware, and Personality Options Assemblies from the Inside Rear Panel View Figure 2-4 Assemblies from the Inside Rear Panel View Agilent E8663B Analog Signal Generator Service Guide...
Replaceable Parts Assemblies and Hardware Options Assemblies from a Microcircuit Deck View Figure 2-5 Assemblies from a Microcircuit Deck View A25B Agilent E8663B Analog Signal Generator Service Guide...
Replaceable Parts Frequency, Hardware, and Personality Options Assemblies from a Bottom View Figure 2-6 Assemblies from a Bottom View Agilent E8663B Analog Signal Generator Service Guide...
Replaceable Parts Assemblies and Hardware Options A18BT1 Lithium Battery Figure 2-7 A18BT1 Lithium Battery A18BT1 2-10 Agilent E8663B Analog Signal Generator Service Guide...
Page 89
Semi- Rigid Cable, A24 20 GHz A30J2 E8251- 20093 Highband Coupler IN Semi- Rigid Cable, A24 20 GHz AT1 115 dB Mechanical Attenuator E8251- 20094 Highband Coupler output input A37J3 A30J2 E8251- 20219 2-12 Agilent E8663B Analog Signal Generator Service Guide...
Page 90
E8251- 20333 (with Option UNW) Filter) (OUT) W121 A29 20 GHz Doubler AT1 115 dB Mechanical Attenuator E8251- 20026 (IN) This cable is not replaceable by itself. You must order the corresponding assembly. 2-13 Agilent E8663B Analog Signal Generator Service Guide...
“Type- N RF Output Connector” on page 2- 26 • “Main Chassis with Inside Cover and Right- Side Support Wall” on page 2- 27 • “Outer and Inner Instrument Covers and Associated Parts” on page 2- 28 2-14 Agilent E8663B Analog Signal Generator Service Guide...
Screw- Machine M3.0 x 08 FL- TX 0515- 1227 Cable Retainer Clip 1400- 1439 Sub Panel E8251- 00009 Screw- Machine M3.0 x 06 CW- PN- TX 0515- 0372 Screw- Machine M3.0 x 08 FL- TX 0515- 1035 2-16 Agilent E8663B Analog Signal Generator Service Guide...
Page 94
Replaceable Parts Frequency, Hardware, and Personality Options Figure 2-9 Disassembled Front Panel View 2-17 Agilent E8663B Analog Signal Generator Service Guide...
Standoff Screw- Hex .327 6- 32 (GPIB interface) 0380- 0644 Split Lock Washer .194ID 10 (GPIB interface) 2190- 0577 Screw- Machine M3.0 8 FL- TX (line module) 0515- 1035 Right Rear Strut E8251- 00006 Left Rear Strut E8251- 00030 2-18 Agilent E8663B Analog Signal Generator Service Guide...
Page 96
Replaceable Parts Frequency, Hardware, and Personality Options Figure 2-10 Disassembled Rear Panel View 2-19 Agilent E8663B Analog Signal Generator Service Guide...
Microcircuit Interface Deck Top View Table 2-8 Microcircuit Interface Deck Top View Item No. Description Part Number Screw- Machine M3.0 x 14 CW- PN- TX 0515- 0665 Figure 2-14 Microcircuit Interface Deck Top View 2-23 Agilent E8663B Analog Signal Generator Service Guide...
Screw- Machine M30. x 20 CW- PN- TX (A30 Modulator Filter with Standard or 0515- 1410 High Power to MID) Screw- Machine M30. x 20 CW- PN- TX (A29 20 GHz Doubler to MID) 0515- 1410 2-24 Agilent E8663B Analog Signal Generator Service Guide...
Page 102
Replaceable Parts Frequency, Hardware, and Personality Options Figure 2-15 Main Chassis and MID Bottom View (3 places) (15 places) (2 places) (3 places) (2 places) 2-25 Agilent E8663B Analog Signal Generator Service Guide...
Screw- Machine M3.0 x 08 FL- TX (microcircuit interface deck to right- side support wall) 0515- 1035 Right- Side Support Wall E8251- 00014 Microcircuit Interface Deck E8251- 00004 Main Chassis E8251- 60354 Figure 2-17 Main Chassis with Inside Cover and Right-Side Support Wall 2-27 Agilent E8663B Analog Signal Generator Service Guide...
E8663- 90004 Programming Guide E8663- 90005 SCPI Command Reference E8663- 90006 Service Guide E8663- 90009 Specifications/Data Sheet 5989- 4866EN NOTE The E8663B documentation can also be found on the web at http://www.agilent.com/find/e8663b. 2-29 Agilent E8663B Analog Signal Generator Service Guide...
(screws) external flathead (screws) feet hexagonal GPIB general purpose interface bus hexagonal recess (screws) in- phase low frequency meters or metric hardware patch lock (screws) or printed circuit 2-30 Agilent E8663B Analog Signal Generator Service Guide...
Page 108
Replaceable Parts Frequency, Hardware, and Personality Options Abbreviation Definition panhead (screws) quadrature quantity radio frequency reverse power protection subminiature type- A TORX recess (screws) 2-31 Agilent E8663B Analog Signal Generator Service Guide...
Page 109
Replaceable Parts Assemblies and Hardware Options 2-32 Agilent E8663B Analog Signal Generator Service Guide...
“A29 20 GHz Doubler” on page 3- 46 • “A30 Modulation Filter” on page 3- 48 • “A31 Motherboard” on page 3- 50 • “A32 10 MHz Crystal Oscillator” on page 3- 52 Agilent E8663B Analog Signal Generator Service Guide...
Page 111
Assembly Replacement Assemblies That You Can Replace • “Rear Panel” on page 3- 54 • “AT1 115 dB Attenuator” on page 3- 56 • “B1 Fan” on page 3- 58 Agilent E8663B Analog Signal Generator Service Guide...
6. Tilt the signal generator forward and slide the outer cover (6) back to remove it from the frame. Replacement Procedure • Reverse the order of the removal procedures. • Torque all T- 15 screws to 21 in- lbs. • Torque all T- 20 screws to 21 in- lbs. Agilent E8663B Analog Signal Generator Service Guide...
Page 113
Assembly Replacement Outer Instrument Cover Figure 3-1 Outer Instrument Cover Removal Agilent E8663B Analog Signal Generator Service Guide...
4. Using the T- 10 driver, remove the twelve screws (1) from the inner cover (2). 5. Remove the inner cover. Replacement Procedure • Reverse the order of the removal procedures. • Torque all T- 10 screws to 9 in- lbs. Figure 3-2 Inner Instrument Cover Removal Agilent E8663B Analog Signal Generator Service Guide...
9. Disconnect the A1W1 display cable from the A31 Motherboard P13. 10. Disconnect the A4W1 inverter cable from the A31 Motherboard P12. 11. Disconnect the W9 ribbon cable from the A31 Motherboard P11. 12. Remove the front panel. Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all T- 15 screws to 21 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Figure 3-3 Front Panel Removal Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedures. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Agilent E8663B Analog Signal Generator Service Guide...
Page 118
Assembly Replacement A1 Keyboard Figure 3-4 A1 Keyboard Removal Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-10 Agilent E8663B Analog Signal Generator Service Guide...
Page 120
Assembly Replacement A2 Display Figure 3-5 A2 Display Removal 3-11 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-12 Agilent E8663B Analog Signal Generator Service Guide...
Page 122
Assembly Replacement A2DS1 Display Backlight Figure 3-6 A2DS1 Display Backlight Removal 3-13 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-14 Agilent E8663B Analog Signal Generator Service Guide...
Page 124
Assembly Replacement A3 Power Switch Figure 3-7 A3 Power Switch Removal 3-15 Agilent E8663B Analog Signal Generator Service Guide...
5. Remove the A4 Inverter by carefully pulling it away from the front panel. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Figure 3-8 A4 Inverter Removal 3-16 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Figure 3-9 A5 Sampler, A6 Frac-N, A7 Reference, and A8 Output 3-17 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Figure 3-10 A9 YIG Driver W54/A9J9 W12/A9J5 W10/A9J3 W11/A9J4 3-18 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. Figure 3-11 A10 ALC W4/A10J4 W14/A10J5 W13/A10J201 W15/A10J3 W55/A10J304 3-19 Agilent E8663B Analog Signal Generator Service Guide...
Lift the A11 Pulse/Analog Modulation Generator out of the slot. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-20 Agilent E8663B Analog Signal Generator Service Guide...
Using the T- 10 driver, remove the sheet metal from A12 Pulse Modulator Board. Replacement Procedure • Reverse the order of the removal procedure. • Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-22 Agilent E8663B Analog Signal Generator Service Guide...
Page 132
Assembly Replacement A12 Pulse Modulator Board Figure 3-13 A12 Pulse Modulator Board 3-23 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Send command "DIAG:FILE:REST", this copies data to the CPU board. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-24 Agilent E8663B Analog Signal Generator Service Guide...
Page 134
Assembly Replacement A18 CPU Figure 3-14 A18 CPU W66/A18P2 W18/A18J1 W19/A18J2 3-25 Agilent E8663B Analog Signal Generator Service Guide...
1. To install the battery, the positive side is aligned with the positive sign on the A18’s battery clip. 2. Reverse the remaining steps of the removal procedure. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-26 Agilent E8663B Analog Signal Generator Service Guide...
Page 136
Assembly Replacement A18BT1 Figure 3-15 A18BT1 W66/A18P2 W19/A18J2 W18/A18J1 3-27 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-28 Agilent E8663B Analog Signal Generator Service Guide...
Page 138
Assembly Replacement A19 Power Supply Figure 3-16 A19 Power Supply Removal A22W2 A22W1 W66/A18J2 W18/A18J1 3-29 Agilent E8663B Analog Signal Generator Service Guide...
3. Torque all hex- nuts to 9 in- lbs. 4. Torque all BNC nuts to 21 in- lbs. 5. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-30 Agilent E8663B Analog Signal Generator Service Guide...
Page 140
Assembly Replacement A21 Rear-Panel Interface Figure 3-17 A21 Rear-Panel Interface 3-31 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all nuts to 9 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-32 Agilent E8663B Analog Signal Generator Service Guide...
Page 142
Assembly Replacement A22 Line Module Figure 3-18 A22 Line Module Removal 3-33 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-34 Agilent E8663B Analog Signal Generator Service Guide...
Page 144
Assembly Replacement A23 Lowband Coupler/Detector Figure 3-19 A23 Lowband Coupler/Detector 3-35 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-36 Agilent E8663B Analog Signal Generator Service Guide...
Page 146
Assembly Replacement A24 20 GHz Highband Coupler Figure 3-20 A24 20 GHz Highband Coupler A25B 3-37 Agilent E8663B Analog Signal Generator Service Guide...
6. Remove the A25 20 GHz Highband Detector. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-38 Agilent E8663B Analog Signal Generator Service Guide...
Page 148
Assembly Replacement A25 20 GHz Highband Detector Figure 3-21 A25 20 GHz Highband Detector 3-39 Agilent E8663B Analog Signal Generator Service Guide...
• Reverse the order of the removal procedure. • Torque all T- 10 screws to 9 in- lbs. • Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-40 Agilent E8663B Analog Signal Generator Service Guide...
Page 150
Assembly Replacement A25B Highband Detector Bias Board Figure 3-22 A25B Highband Detector Bias Board A25BW1 A25B 3-41 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-42 Agilent E8663B Analog Signal Generator Service Guide...
Page 152
Assembly Replacement A26 MID (Microcircuit Interface Deck) Figure 3-23 A26 MID W113/A26J34 W20/A26J20 W36/A26J13 W28A26J15 W69/A26J14 W35/A26J10 W19/A26J30 W34/A26J31 W33/A26J32 3-43 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-44 Agilent E8663B Analog Signal Generator Service Guide...
Page 154
Assembly Replacement A28 YIG Oscillator Figure 3-24 A28 YIG Oscillator 3-45 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-46 Agilent E8663B Analog Signal Generator Service Guide...
Page 156
Assembly Replacement A29 20 GHz Doubler Figure 3-25 A29 20 GHz Doubler W41/A29J2 W33/A29J10 W39/A29J4 3-47 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-48 Agilent E8663B Analog Signal Generator Service Guide...
Page 158
Assembly Replacement A30 Modulation Filter Figure 3-26 A30 Modulator Filter W41/A30J1 W108/A30J6 W17/A30J7 W34/A30J10 W27/A30J3 W107/A30J2 3-49 Agilent E8663B Analog Signal Generator Service Guide...
3. Torque all T- 10 screws to 9 in- lbs. 4. Torque all RF connectors to 8 in- lbs. 5. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-50 Agilent E8663B Analog Signal Generator Service Guide...
Page 160
Assembly Replacement A31 Motherboard Figure 3-27 A31 Motherboard W23/A31P241 W35/A31P201 W40/A29J3 W39/A29J4 W21/A31J1043 W22/A31J1042 W47/A31J1041 W11/A31J1021 W10/A31J1031 W54/A31J1032 3-51 Agilent E8663B Analog Signal Generator Service Guide...
1. Reverse the order of the removal procedure. 2. Torque all T- 10 screws to 9 in- lbs. 3. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-52 Agilent E8663B Analog Signal Generator Service Guide...
Page 162
Assembly Replacement A32 10 MHz Crystal Oscillator Figure 3-28 A32 Crystal Oscillator 3-53 Agilent E8663B Analog Signal Generator Service Guide...
• Torque all T- 10 screws to 9 in- lbs. • Torque all RF connectors to 8 in- lbs. • Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-54 Agilent E8663B Analog Signal Generator Service Guide...
Page 164
Assembly Replacement Rear Panel Figure 3-29 Rear Panel Removal A22W2 A22W1 W18/A18J1 3-55 Agilent E8663B Analog Signal Generator Service Guide...
2. Torque all T- 10 screws to 9 in- lbs. 3. Torque all RF connectors to 8 in- lbs. 4. Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-56 Agilent E8663B Analog Signal Generator Service Guide...
Page 166
Assembly Replacement AT1 115 dB Attenuator Figure 3-30 AT1 115 dB Attenuator AT1W1/A26J13 3-57 Agilent E8663B Analog Signal Generator Service Guide...
• Torque all T- 10 screws to 9 in- lbs. • Torque all RF connectors to 8 in- lbs. • Perform the post- repair adjustments and performance tests that pertain to this removal procedure. 3-58 Agilent E8663B Analog Signal Generator Service Guide...
Page 168
Assembly Replacement B1 Fan Figure 3-31 B1 Fan 3-59 Agilent E8663B Analog Signal Generator Service Guide...
Page 169
Assembly Replacement B1 Fan 3-60 Agilent E8663B Analog Signal Generator Service Guide...
Page 170
Power Flatness Calibration ✓ ✓ ALC Modulator Calibration ✓ ✓ ✓ AM Gain Calibration ✓ ✓ ✓ Gain Adjustment Bypass Calibration (Frequencies ≤3.2 GHz) ✓ ✓ Power Clamp ✓ ✓ ✓ Pulse Width Calibration Agilent E8663B Analog Signal Generator Service Guide...
Upconverter Gain Adjustment Calibration ✓ ✓ ✓ Power Flatness Calibration ✓ ALC Modulator Calibration ✓ AM Gain Calibration ✓ Pulse Width Calibration ✓ Lowband Attenuator Calibration (<3.2GHz) ✓ High Band Attenuator Calibration (>3.2GHz) Agilent E8663B Analog Signal Generator Service Guide...
Page 172
✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Self Tests ✓ ✓ ✓ ✓ ✓ ✓ Maximum Leveled Output Power ✓ ✓ ✓ ✓ ✓ ✓ Power Level Accuracy ✓ Internal FM Frequency Response ✓ External Phase Modulation Frequency Response Agilent E8663B Analog Signal Generator Service Guide...
Page 173
Internal Pulse Modulation Rise/Fall Time ✓ External Pulse Modulation ON/OFF Ratio ✓ Harmonic Spurious Sub- Harmonic Spurious ✓ ✓ ✓ Non- Harmonic Spurious ✓ Single- Sideband Phase Noise Optional Performance Test ✓ Timebase Aging Rate Agilent E8663B Analog Signal Generator Service Guide...
This product must be used in a normal condition (in which all means for protection are intact) only. CAUTION Many of the assemblies in this instrument are very susceptible to damage from electrostatic discharge (ESD). Perform service procedures only at a static- safe workstation and wear a grounding strap. Agilent E8663B Analog Signal Generator Service Guide...
Page 175
Lithium Battery Disposal When the battery on the A14 CPU is exhausted and/or ready for disposal, dispose of it according to your country’s requirements. You can return the battery to your nearest Agilent Technologies Sales and Service office for disposal, if required.
Page 176
2-40 A6 Frac- N, 2-17 lowband, replacement, 2-34 A7 Reference, 2-17 device gains and losses, 1-64 A8 Output, 2-17 display backlight, replacement, 2-12 A9 YIG driver, 2-18 2-10 display, replacement, Agilent E8663B Analog Signal Generator Service Guide...
Page 177
(MID), replacement, 2-42 2-48 modulator filter, replacement, safety, general considerations, motherboard 2-17 sampler, replacement, 1-14 connector locations, SCPI 1-16 LED locations, reference content, ii-ix test points, 1-11 self- tests motherboard, replacement, 2-50 2xx, 1-18 Agilent E8663B Analog Signal Generator Service Guide...
Page 178
1-46 spurious non- harmonic emissions, 1-47 unlevels, 1-37 1-35 unlocks, where to begin, unlevels, troubleshooting, 1-37 unlocks, troubleshooting, 1-35 user documentation content, ii-ix voltages, supply 1-18 A5 Sampler, Agilent E8663B Analog Signal Generator Service Guide...
Page 179
Agilent E8663B Analog Signal Generator Service Guide...