Key Reference
F
Freq
Supported
This softkey enables you to set the signal generator to sweep frequency data only. The
amplitude is set at a constant level determined by the
a sweep containing both amplitude and frequency information and still choose to sweep
frequency only.
Preset
Range
Location
Remarks
Freq Multiplier
Supported
This softkey enables you to multiply the frequency shown on the display without
changing the frequency output at the RF OUTPUT connector (simulating the frequency
at the output of a harmonic multiplier).
For any multiplier other than 1, the MULT indicator is shown in the frequency area of the
display.
Preset
Range
Location
Remarks
152
All
N/A
N/A
Page 245
None
All
1
−100 to +1000 (excluding 0)
Page 239
The displayed frequency value is equal to the actual frequency, minus
the reference frequency, times the multiplier, plus the offset frequency.
hardkey. You can define
Amplitude
Chapter 5