Configuring the LF Output (PSG-A Series Only)

The PSG-A Series signal generator has a low frequency (LF) output. The LF output's source
can be switched between an internal modulation source (internal monitor 1 or 2) or an
internal function generator (function generator 1 or 2).
Using internal monitor 1 or 2 as the LF output source, the LF output provides a replica of the
signal from either internal source (1 or 2) that is being used to modulate the RF output. The
specific modulation parameters for this signal are configured through the AM, FM, or ΦM
menus.
Using function generator 1 or 2 as the LF output source, the function generator section of the
internal modulation source (1 or 2) drives the LF output directly. Frequency and waveform
are configured from the LF output menu, not through the AM, FM, or ΦM menus. You can
select the waveform shape from the following choices:
Sine
sinewave with adjustable amplitude and frequency
Dual-Sine
dual sinewaves with individually adjustable frequencies and a percent-of-
peak-amplitude setting for the second tone (available from function
generator 1 only)
Swept-Sine
a swept sinewave with adjustable start and stop frequencies, sweep time,
and sweep trigger settings (available from function generator 1 only)
Triangle
triangle wave with adjustable amplitude and frequency
Ramp
ramp with adjustable amplitude and frequency
Square
square wave with adjustable amplitude and frequency
Noise
noise, in a uniform or Gaussian distribution, with adjustable amplitude
generated as a peak-to-peak value (RMS value is approximately 80% of the
displayed value)
DC
direct current with adjustable amplitude
NOTE
The
LF Out Off On
LF output source is set to function generator 1 or 2. The
controls the operating state of the LF OUTPUT connector when the LF output
source is set to internal monitor 1 or 2.
The
RF On/Off
Chapter 2
softkey controls the operating state of the LF output when the
softkey does not apply to the LF OUTPUT connector.
Configuring the LF Output (PSG-A Series Only)
Mod On/Off
Fundamental Operation
softkey
41