2. Width:
3. Height:
4. Before operating the instrument, verify
the ambient temperature:
5. Verify the operating humidity:
6. Verify the operating altitude:
7. Maximum input voltage, <4 GHz models:
50 Ω
1 MΩ
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DPO5000 Series User Manual
17.96 inches (456 mm)
10.9 inches (277 mm)
5 °C to +45 °C (+41 °F to +113 °F)
8% to 80% relative humidity with
a maximum wet-bulb temperature
of +29 °C (+84 °F) at or below
+45 °C (+113 °F), noncondensing
Upper limit derated to 30% relative
humidity at +45 °C (+113 °F)
<4 GHz models: 3,000 m
(9,843 feet)
5 V
, with peaks ≤±24 V.
RMS
150 V, derate at 20 dB/decade to 9 V
input voltage at the BNC, between center conductor and ground is
400 V peak. The RMS voltage is limited to <150 V for arbitrary waveshapes
including DC. The maximum pulse width for impulses with peaks over
150 V is 50 μs. Example: At 0 V to 400 V peak, rectangular wave, the duty
factor is 14%. The maximum transient withstand voltage is ±800 V peak.
Install Your Instrument
17.75 inches (451 mm)
11.48 inches (292 mm)
5 °C to +45 °C (+41 °F to +113 °F)
8% to 80% relative humidity at up
to +32 °C (+90 °F)
5% to 45% relative humidity above
+32 °C (+90 °F) up to +45 °C
(+113 °F), noncondensing, and
is limited by a maximum wet-bulb
temperature of +29.4 °C (+85 °F)
(derates relative humidity to 32%
at +45 °C (+113 °F))
≥4 GHz models: 3,000 m
(9,843 feet), derate maximum
operating temperature by 1 °C
per 300 meters (984.25 feet)
above 1500 meters (4921.25 feet)
altitude.
above 200 kHz. The maximum
RMS